Mon, 29 Apr CM Problem Sheet 4 Electroanalysis_918840047 - Saved Home Insert Draw Layout Review View I X2 More Formatting 日 65% 2. A potassium ion selective electrode used for monitoring blood potassium levels during heart surgery was reported as being defective during routine calibration. Using the calibration data below can you: (i) Confirm the report (ii) Explain your reasoning Standard Reading 1 Reading 2 Reading 3 1 mM -15.0 -15.9 -14.9 2 mM -4.1 -4.3 -4.8 4 mM 7.1 7.4 7.6 6 mM 13.8 14.5 14.2 8 mM 20.3 19.8 20 10 mM 23.4 23.9 24.1 KC fruit juices by a differential pulse
Mon, 29 Apr CM Problem Sheet 4 Electroanalysis_918840047 - Saved Home Insert Draw Layout Review View I X2 More Formatting 日 65% 2. A potassium ion selective electrode used for monitoring blood potassium levels during heart surgery was reported as being defective during routine calibration. Using the calibration data below can you: (i) Confirm the report (ii) Explain your reasoning Standard Reading 1 Reading 2 Reading 3 1 mM -15.0 -15.9 -14.9 2 mM -4.1 -4.3 -4.8 4 mM 7.1 7.4 7.6 6 mM 13.8 14.5 14.2 8 mM 20.3 19.8 20 10 mM 23.4 23.9 24.1 KC fruit juices by a differential pulse
Chapter11: Dynamic Electrochemistry
Section: Chapter Questions
Problem 10P
Related questions
Question
![Mon, 29 Apr CM
Problem Sheet 4 Electroanalysis_918840047 - Saved
Home
Insert
Draw Layout
Review View
I
X2 More Formatting
日
65%
2. A potassium ion selective electrode used for monitoring blood potassium levels during heart
surgery was reported as being defective during routine calibration. Using the calibration
data below can you:
(i)
Confirm the report
(ii) Explain your reasoning
Standard
Reading 1
Reading 2
Reading 3
1 mM
-15.0
-15.9
-14.9
2 mM
-4.1
-4.3
-4.8
4 mM
7.1
7.4
7.6
6 mM
13.8
14.5
14.2
8 mM
20.3
19.8
20
10 mM
23.4
23.9
24.1
KC
fruit juices by a differential pulse](/v2/_next/image?url=https%3A%2F%2Fcontent.bartleby.com%2Fqna-images%2Fquestion%2Fe8d3d618-55be-44c2-9091-0170276b597d%2Fb7f50def-fbe9-4da7-84eb-f1c081b9662a%2F1evjivi_processed.jpeg&w=3840&q=75)
Transcribed Image Text:Mon, 29 Apr CM
Problem Sheet 4 Electroanalysis_918840047 - Saved
Home
Insert
Draw Layout
Review View
I
X2 More Formatting
日
65%
2. A potassium ion selective electrode used for monitoring blood potassium levels during heart
surgery was reported as being defective during routine calibration. Using the calibration
data below can you:
(i)
Confirm the report
(ii) Explain your reasoning
Standard
Reading 1
Reading 2
Reading 3
1 mM
-15.0
-15.9
-14.9
2 mM
-4.1
-4.3
-4.8
4 mM
7.1
7.4
7.6
6 mM
13.8
14.5
14.2
8 mM
20.3
19.8
20
10 mM
23.4
23.9
24.1
KC
fruit juices by a differential pulse
Expert Solution
![](/static/compass_v2/shared-icons/check-mark.png)
This question has been solved!
Explore an expertly crafted, step-by-step solution for a thorough understanding of key concepts.
Step by step
Solved in 2 steps with 1 images
![Blurred answer](/static/compass_v2/solution-images/blurred-answer.jpg)
Recommended textbooks for you
![Principles of Instrumental Analysis](https://www.bartleby.com/isbn_cover_images/9781305577213/9781305577213_smallCoverImage.gif)
Principles of Instrumental Analysis
Chemistry
ISBN:
9781305577213
Author:
Douglas A. Skoog, F. James Holler, Stanley R. Crouch
Publisher:
Cengage Learning
![Principles of Instrumental Analysis](https://www.bartleby.com/isbn_cover_images/9781305577213/9781305577213_smallCoverImage.gif)
Principles of Instrumental Analysis
Chemistry
ISBN:
9781305577213
Author:
Douglas A. Skoog, F. James Holler, Stanley R. Crouch
Publisher:
Cengage Learning
![Chemistry: Principles and Practice](https://www.bartleby.com/isbn_cover_images/9780534420123/9780534420123_smallCoverImage.gif)
Chemistry: Principles and Practice
Chemistry
ISBN:
9780534420123
Author:
Daniel L. Reger, Scott R. Goode, David W. Ball, Edward Mercer
Publisher:
Cengage Learning