Interpretation: The types of mass spectrometers that are used in ICPMS needs to be determined. The difference between them needs to be explained.
Concept introduction:Mass spectrometer is an instrument which differentiates particles according to the mass and charge. It has
Answer to Problem 11.1QAP
Based on the techniques used in respective mass spectroscopy following are the different types of spectrometers:
- Quadrupole mass filter
- Time-of-flight
- Sector instruments
- Ion traps:
- Three-dimensional quadrupole
- Cylindrical
- Linear and Orbitrap
Explanation of Solution
Quadrupole analyzers utilize the wavering electric field to specifically balance out or destabilize the ways of particles going in a radio frequency(RF). quadrupole field which is framed between four parallel poles. Particles in a specific scope of mass/charge proportion are gone through the framework whenever, yet varieties to the possibilities on the bars permits a wide scope of m/z esteems to be cleared rapidly, either consistently or in a progression of characterized turns.
In the time-of-flight (TOF) method, electric field is adopted to accelerate ions via same potential. The time taken by ions to reach the detector is then observed.
If all the ions are having the same charge, then the kinetic energy tend to be similar, and the ions' velocity will depend only on the mass of the ions and it can be seen that the ions which are having a lower mass will be reaching the detector earlier than those having higher mass.
A segment field strategy or (A sector field) pursues a system that incorporates a static electric and additionally magnet attractive field to impact the speed or velocity of the particles. It twists the directions of the particles. As indicated by their mass-to-charge proportions, it very well may be found that there is more redirection of the more charged and quicker moving. This technique is utilized to choose a scope of m/z or to screen through a scope of m/z to arrange the particles present in a given issue.
Ion traps:
Three-dimensional quadrupole ion trap- This method is based on the principle that is used in the quadrupole method but there is a difference. The ions are trapped and ejected one by one. They are trapped in a quadrupole RF field. Which is defined by a space by a ring electrode and between two end cap electrodes. The sample may be internally or externally ionized inside it, in this case the ions are often introduced by an aperture in an end cap electrode.
Cylindrical particle trap-This technique is a kind of quadrupole method.Here, the terminals i.e the electrodes are framed from level rings, not from hyperbolic formed electrodes. The system permits itself for scaling down as the size of the trap is diminished; the state of the electric field which is at the focal point of the trap frames a shape like a hyperbolic trap.
Straight quadrupole particle trap - This strategy additionally shares similitude to a quadrupole particle trap, yet the difference is that it can trap particles in a two-dimensional quadrupole field, it doesn't utilize three-dimensional quadrupole field as utilized in a 3D quadrupole.
Orbitrap instruments methods have trapping technique for ions in anelectrostatically circular path around a central, spindle shaped electrode.
Hence, this can be concluded that types mass spectrometers are based on the techniques used in respective mass spectroscopy.
Want to see more full solutions like this?
Chapter 11 Solutions
Principles of Instrumental Analysis
- Mass spectrometry is an extremely versatile detection system for GC. However, interfacing an HPLC system to a mass spectrometer is a much more difficult task. Describe the major reasons why it is more difficult to combine HPLC with mass spectrometry than it is to combine GC with mass spectrometry.arrow_forwardWhen LC is coupled with mass spectrometry (MS), what are the advantages (compared to LC only) in terms of analyte identification and quantitation in the sample?arrow_forwardBriefly explain the principles of APCI mass spectrometryarrow_forward
- Which of the following statements is incorrect? 1.)A conventional mass spectrometer does not employ a spectrophotometric detector. 2.)Conventional mass spectrometry does not always require high purity samples. 3.)A mass spectrum shows no signals due to uncharged species. 4.)A conventional mass spectrometer uses high energy UV radiation.arrow_forwardHow is light and energy used in gas chromatography- mass spectrometry (GC-MS)?arrow_forwarda) Explain the fundamentals of gas chromatographyb) Explain the fundamentals of mass spectrometryarrow_forward
- If a TOF mass spectrometer needs to scan an m/z of 600, what is the speed of this ion if it is accelerated through 4000 volts? How long will this ion (at the speed calculated) take to drift 2.00 m through the spectrometer? How many mass spectra can be taken each second if a new cycle begins each time this ion reaches the detector?arrow_forwardIs light used in gas chromatography- mass spectrometry? If so, how is it used?arrow_forwardThe LC-MS (mass spectrometry) is set to positive mode (can only see positively charged ions). What is the other routine mode which the instrument can be set in and when would this mode be used and why?arrow_forward
- Determine the isotopic peak scheme for [ReCl]+ fragment in the mass spectrum? 185Re: %37 ; 187Re: %63; 35Cl: %76; 37Cl: %24arrow_forwardA technique called “GCMS” is often used instead of simple GC analysis. This technique entails passing a sample through a gas chromatography instrument and replacing the standard GC detector with a mass spectrometer (the “MS” part). What advantages do you think GCMS has over GC as well as simple MS?arrow_forward
- Principles of Instrumental AnalysisChemistryISBN:9781305577213Author:Douglas A. Skoog, F. James Holler, Stanley R. CrouchPublisher:Cengage Learning