EBK PRINCIPLES OF INSTRUMENTAL ANALYSIS
EBK PRINCIPLES OF INSTRUMENTAL ANALYSIS
7th Edition
ISBN: 8220103611749
Author: Crouch
Publisher: YUZU
Question
Book Icon
Chapter 15, Problem 15.16QAP
Interpretation Introduction

(a)

Interpretation:

The chemistry of the analytical method should be explained.

Concept introduction:

Quenching is the process of nonradiative energy transfer from an excited species to another molecule. This requires a contact between the excited species and a quenching agent. The rate of quenching is higher when the quencher concentration is high, because of higher number of collisions during the lifetime of the excited state.

Expert Solution
Check Mark

Answer to Problem 15.16QAP

Al-acid alizarin garnet R acts as a quencher. The decrease in fluorescence is due to absorbing emission energy from F- by Al-acid alizarin garnet R complex.

Explanation of Solution

In this experiment standard addition technique has been used. Four water samples with same volume are taken and they have been spiked with standard F-. Different volumes of standard F- solution are added to the water samples prepared. Then excess, but same volumes of Al-acid alizarin garnet R are added to the four solutions. This Al-acid alizarin garnet R acts as a quencher. The complex absorbs emission energy from F-. So, instead of producing fluorescence from emission energy, it is quenched by the complex. So, the fluorescence intensity decreases.

Interpretation Introduction

(b)

Interpretation:

A plot of data should be constructed.

Concept introduction:

Standard addition method is used to analyze complex samples which matrix effect is significant. One common approach is standard addition technique is spiking of sample. In this method, one or more increments of a standard solution is added to the several aliquots of sample containing same volume. Each solution is then diluted to a fixed volume and experiment is proceeded.

Expert Solution
Check Mark

Answer to Problem 15.16QAP

EBK PRINCIPLES OF INSTRUMENTAL ANALYSIS, Chapter 15, Problem 15.16QAP , additional homework tip  1

Explanation of Solution

Meter reading is the dependent variable and the volume of standard added is the independent variable.

EBK PRINCIPLES OF INSTRUMENTAL ANALYSIS, Chapter 15, Problem 15.16QAP , additional homework tip  2

Interpretation Introduction

(c)

Interpretation:

A relationship for multiple standard additions should be derived. An equation for the unknown concentration should be obtained in terms of the slope and the intercept of the standard addition plot.

Concept introduction:

Several aliquots with volume Vx of an unknown solution which have a concentration Cx are measured into volumetric flasks with volume Vt and then a standard solution with concentration Csis added in variable volumes of Vs. Then other reagents that aid for the detection are added and each volumetric flask is topped up to the maximum volume. Then instrumental measurements are made on each solution. If the instrumental response is proportional to the concentration of analyte, following equation can be derived.

S=kVscsVt+kVxcxVt  (1)

k is a proportionality constant. According to above equation the instrumental signal versus Vs is a plot in the form of S=mVs+b, where,

m=kcsVt

b=kVxcxVt

Concentration of unknown then can be obtained by following equation.

cx=bcsmVx   (2)

Expert Solution
Check Mark

Answer to Problem 15.16QAP

S=kVxcxVtkVscsVt

cx=bcsmVx

Explanation of Solution

The fluorescence decreases with the increasing amounts of F- standard. The control sample has the highest fluorescence meter reading. So the equation 1 can be rewrite as

The meter reading versus the volume of standard added has a linear relationship. Above equation can be written in the form of S=mVs+b

S=kVscsVt+kVxcxVt where,

m=kcsVtb=kVxcxVt

An equation for the concentration of unknown can be derived by taking the ratio of intercept and the slope of fluorescence Vs Vs graph.

bm=kVxcx/Vtkcs/Vt=Vxcxcscx=bcsmVx

Interpretation Introduction

(d)

Interpretation:

The equation for the line representing the decrease in fluorescence relative to volume of standard should be determined.

Concept introduction:

In the least squares method, in order to find the slope and the intercept the three quantities Syy, Sxx and Sxy should be determined.

Syy=(yiy¯)2Sxx=(xix¯)2Sxy=(xix¯)(yiy¯)

The slope of the line, m=SxySxx

The intercept, b=y¯mx¯

y¯ - mean of y values

x¯ - mean of x values

Expert Solution
Check Mark

Answer to Problem 15.16QAP

y=13.22x+68.23

Explanation of Solution

The three quantities Syy, Sxx and Sxy can be easily calculated using a spreadsheet and then the slope and the intercept can be calculated using the given equations.

Vs, mL Meter reading (xix¯) (yiy¯) (yiy¯)2 (xix¯)2 (xix¯)(yiy¯)
0 68.2 -1.5 19.8 392.04 2.25 -29.7
1 55.3 -0.5 6.9 47.61 0.25 -3.45
2 41.3 0.5 -7.1 50.41 0.25 -3.55
3 28.8 1.5 -19.6 384.16 2.25 -29.4
 Sum        874.22  5 -66.1
y¯ 48.4        
x¯ 1.5        

m=SxySxx=66.15=13.22

b=y¯mx¯b=48.4(13.22×1.5)b=68.23

So the equation for the line for the standard addition plot is y=13.22x+68.23

Interpretation Introduction

(e)

Interpretation:

The standard deviation of the slope and the intercept should be determined.

Concept introduction:

Standard deviation about regression, sr=Syym2SxxN2

N − number of points used.

The standard deviation of the slope, sm=sr2Sxx

The standard deviation of the intercept, sb=srxi2Nxi2(xi)2

Expert Solution
Check Mark

Answer to Problem 15.16QAP

Standard deviation of the slope = 0.194

Standard deviation of the intercept = 0.837

Explanation of Solution

in order to calculate the standard deviation of the slope or intercept, first standard deviation about regression should be determined.

sr=874.22(13.22)2×542sr=0.435

Standard deviation of the slope, sm=(0.435)25=0.194

Standard deviation of the intercept, sb=sr144×14(6)2=0.837

Interpretation Introduction

(f)

Interpretation:

The concentration of F- in the sample in parts per billion should be determine.

Concept introduction:

cx=bcsmVx

Cx − concentration of the unknown solution

Cs − concentration of standard solution

Vx − volume of unknown solution

m − slope of the instrumental signal versus volume of standard added curve

b − intercept of the instrumental signal versus volume of standard added curve

Expert Solution
Check Mark

Answer to Problem 15.16QAP

5.16 ppb

Explanation of Solution

Intercept = 68.23

Slope = -13.22

Concentration of standard = 10.0 ppb

Volume of unknown = 10.00 mL

cx=bcsmVx

cx=68.23×10.0 ppb(13.22)/ mL ×10.00 mLcx=5.16 ppb

Interpretation Introduction

(g)

Interpretation:

The standard deviation of the result in (f) should be determined

Concept introduction:

sv=sym1N+(0y¯)2m2Sxx

Sv − standard deviation in volume

Sy − standard deviation in signal

m −slope

N − number of points

sc=sv(csVx)

Sc − standard deviation in concentration

Expert Solution
Check Mark

Answer to Problem 15.16QAP

sc=3.78 ppb

Explanation of Solution

sv=17.07113.2214+(048.4)2(13.22)2×5sv=3.78

sc=3.78 mL(10.0 ppb10.00 mL)sc=3.78 ppb

Want to see more full solutions like this?

Subscribe now to access step-by-step solutions to millions of textbook problems written by subject matter experts!
Students have asked these similar questions
Exactly 10.00-mL aliquots of a solution containing phenobarbital were measured into 50.00-mL volumetric flasks and made basic with KOH. The following volumes of a standard solution of phenobarbital containing 2.000μg/mL of phenobarbital were then introduced into each flask and the mixture was diluted to volume: 0.000, 0.500, 1.00, 1.50, 2.00 mL. The fluorescence of each of these solutions was measured with a fluorimeter, which gave values of 3.26, 4.80, 6.41, 8.02, 9.56, respectively. a. plot the data.  b. derive a least squares equation for the data plotted in (a). c. find the concentration of phenobarbital from the equation in (b). d. calculate a standard deviation for the concentration obtained in (c).
Seawater was spiked with phosphate to achieve a quantifiable range of phosphate concentrations by spectroscopy. What components of the spectrophotometer or sample preparation could be changed that would improve detection limits?
Mercury(II) forms a 1:1 complex with triphenyltetrazolium chloride that exhibits an absorption maximum at 255 nm. The mercury(II) in a soil sample was extracted into an organic solvent containing an excess of TTC , and the resulting solution was diluted to 100.0 mL in a volumetric flask. Five-milliliter aliquots of the analyte solution were then transferred to six 25-mL volumetric flasks. A standard solution was then prepared that was 5*10^-6 M in . Volumes of the standard solution shown in the table were then pipetted into the volumetric flasks, and each solution was then diluted to 25.00 mL. The absorbance of each solution was measured at 255 nm in 1.00-cm quartz cells. Please solve and explain these questions: A.) Enter the given data pictured above into a spreadsheet and show the correct standard additions plot B.) Determine the slope and the intercept of the line C.)Determine the standard deviation of the slope and the intercept. D.) Calculate the concentration of Hg(II) in the…
Knowledge Booster
Background pattern image
Similar questions
Recommended textbooks for you
Text book image
Principles of Instrumental Analysis
Chemistry
ISBN:9781305577213
Author:Douglas A. Skoog, F. James Holler, Stanley R. Crouch
Publisher:Cengage Learning
Text book image
Fundamentals Of Analytical Chemistry
Chemistry
ISBN:9781285640686
Author:Skoog
Publisher:Cengage
Text book image
Macroscale and Microscale Organic Experiments
Chemistry
ISBN:9781305577190
Author:Kenneth L. Williamson, Katherine M. Masters
Publisher:Brooks Cole