Theory and Design for Mechanical Measurements
6th Edition
ISBN: 9781118881279
Author: Richard S. Figliola, Donald E. Beasley
Publisher: WILEY
expand_more
expand_more
format_list_bulleted
Textbook Question
Chapter 5, Problem 5.36P
A pressure measuring system outputs a voltage that is proportional to pressure. It is calibrated against a transducer standard (certified error: within ±0.5 psi) over its 0-100-psi range with the results given below. The voltage is measured with a voltmeter (instrument error: within ±10 pV; resolution: 1 pV). The engineer intending to use this system estimates that installation effects can cause the indicated pressure to be off by another ±0.5 psi. Estimate the expanded uncertainty at 95% confidence in using the system based on the known information.
£lmVJ: 0.(X)4 0.399 0.771 1.624 2.147 4.121
p[psi]: 0.1 10.2 19.5 40.5 51.2 99.6
Expert Solution & Answer
Want to see the full answer?
Check out a sample textbook solutionStudents have asked these similar questions
The transducer specified in Table 1.1 is chosen to measurea nominal pressure of 500 cm H2O. The ambient temperature is expected to vary between 18 ∘C and 25 ∘C duringtests. Estimate the possible range (magnitude) of each listedelemental error affecting the measured pressure.
How do I calculate the sensitivity error?
The solution that was given:
Sensitivity error(eK)
= (±0.0025)(500 cm H2O)= ± 0.75 cm H2O
= ± 0.00375 V
My Question is, how do you obtain 0.75 since 0.0025 x 500 gives 1.25!
please help. thanks
A displacement transducer has the following specifications:
Linearity error:±0.25% reading
Drift:±0.05%/°C reading
Sensitivity error: ±0.25% reading
Excitation: 10-25 V
Output: dc 0-5Vdc
Range: 0-5 cm
The transducer output is to be indicated on a voltme- ter having a stated accuracy of 0:1% reading with a resolution of 10 μV. The system is to be used at room temperature, which can vary by 10 °C. Estimate an uncertainty in a nominal displacement of 2 cm at the design stage. Assume 95% confidence.
A displacement transducer has the following specifications:
Linearity error: ±0.25% reading
Drift: ±0.05%/°C reading
Sensitivity error: ±0.25% reading
Excitation: 10-25 V dc
Output: 0-5Vdc
Range: 0-5 cm
The transducer output is to be indicated on a voltmeter having a stated accuracy of 0:1% reading with a resolution of 10 μV. The system is to be used at room temperature, which can vary by 10 °C. Assume 95% confidence.
The displacement transducer above is used in measuring the displacement of a body impacted by a mass. Twenty measurements are made, which yield
Determine a best estimate for the mass displacement at 95% probability based on all available information.
Chapter 5 Solutions
Theory and Design for Mechanical Measurements
Ch. 5 - Prob. 5.1PCh. 5 - Explain what is meant by the terms “true value,"...Ch. 5 - SJ An official Olympics timekeeper once stated,...Ch. 5 - A tachometer has an analog display dial graduated...Ch. 5 - An automobile speedometer is graduated in 5-mph...Ch. 5 - An engineer reads the height of a mercury...Ch. 5 - Prob. 5.8PCh. 5 - Two resistors arc to be combined to form an...Ch. 5 - An equipment catalog boasts that a pressure...Ch. 5 - The shear modulus, G, of an alloy can be...
Ch. 5 - Heat transfer from a rod of diameter D immersed in...Ch. 5 - Estimate the design-stage uncertainty in...Ch. 5 - Explain the critical dift'erence(s) between a...Ch. 5 - A displacement transducer has the following...Ch. 5 - Prob. 5.18PCh. 5 - For a thin-walled pressure vessel of diameter D...Ch. 5 - Suppose a measured normal stress contains three...Ch. 5 - Prob. 5.31PCh. 5 - The area of a flat, rectangular parcel of land is...Ch. 5 - Estimate the random standard uncertainty in the...Ch. 5 - Estimate the uncertainty at 95% confidence in the...Ch. 5 - One use for a GPS rangefinder is on the golf...Ch. 5 - A pressure measuring system outputs a voltage that...Ch. 5 - The density of a metal composite is to be...Ch. 5 - A temperature measurement system is calibrated...Ch. 5 - The power usage of a strip heater is to be...Ch. 5 - The power usage of a DC strip heater can be...Ch. 5 - A thermocouple is a temperature sensor made of two...Ch. 5 - Time variations in a signal require that the...Ch. 5 - A hand-held infrared thermometer from a well-...Ch. 5 - The pressure in a large vessel is to be maintained...Ch. 5 - The cooling of a thermometer (c.g., Exs. 3.3 and...Ch. 5 - A J-type thermocouple monitors the temperature of...Ch. 5 - The density of air must be known to within 0.5%....Ch. 5 - In pneumatic conveying, solid particles such as...Ch. 5 - A step test is run to determine the time constant...Ch. 5 - The acceleration of a cart down a plane inclined...Ch. 5 - The heat flux in a reaction is estimated by Q =...Ch. 5 - A comparative test uses the relationship R =...Ch. 5 - A sensitive material is to be contained within a...Ch. 5 - A geometric stress concentration factor, K,, is...Ch. 5 - In Problem 5.9, we assumed that the errors in die...
Knowledge Booster
Learn more about
Need a deep-dive on the concept behind this application? Look no further. Learn more about this topic, mechanical-engineering and related others by exploring similar questions and additional content below.Similar questions
- Using the information in Problem 1.22, estimate the ambient air temperature that could cause frostbite on a calm day on the ski slopes. 1.22 In order to prevent frostbite to skiers on chair lifts, the weather report at most ski areas gives both an air temperature and the wind-chill temperature. The air temperature is measured with a thermometer that is not affected by the wind. However, the rate of heat loss from the skier increases with wind velocity, and the wind-chill temperature is the temperature that would result in the same rate of heat loss in still air as occurs at the measured air temperature with the existing wind. Suppose that the inner temperature of a 3-mm-thick layer of skin with a thermal conductivity of 0.35W/mKis35C and the air temperature is 20C. Under calm ambient conditions the heat transfer coefficient at the outer skin surface is about 20W/m2K (see Table 1.4), but in a 40-mph wind it increases to 75W/m2K. If frostbite occurs when the skin temperature drops to about 10C, do you advise the skier to wear a face mask? What is the skin temperature drop due to the wind?arrow_forwardA displacement transducer has the following specifications: Linearity error: ±0.25% reading Drift: ±0.05%/°C reading Sensitivity error: ±0.25% reading Excitation: 10-25 V dc Output: 0-5Vdc Range: 0-5 cm The transducer output is to be indicated on a voltmeter having a stated accuracy of 0:1% reading with a resolution of 10 μV. The system is to be used at room temperature, which can vary by 10 °C. Assume 95% confidence. The displacement transducer above is used in measuring the displacement of a body impacted by a mass. Twenty measurements are made, which yield X= 17.20mm Sx= 1.70mm Determine a best estimate for the mass displacement at 95% probability based on all available information.arrow_forwardTo measure the density of a stone, a digital scale and a beaker cup filled with water are used. The balance is used to measure the mass of the stone, while the beaker is used to measure the volume of the stone. It is known that the digital balance has a resolution of 0.1 g and that the beaker cup has a resolution of 5 ml. Repeated mass measurements return the following values: 250.0 g; 249.9 g; 250.3 g; 249.6g; 250.1 g; 249.7g; 250.4 g; 250.0 g; 248.8 g; 250.2 g The volume is measured only once. Placing the stone in the beaker, a volume of displaced water equal to 50 ml is measured. Using a k=2.5, check if the density of the stone is compatible or out of line with the value ρ = (4910 ± 65) kg.m-3 (a) none of the above (b) the values are compatible (c) missing data to perform the compatibility test (d) the values are discrepantarrow_forward
- The following resistance of a platinum resistance thermometer were measured at a range of temperature. Determine the measurement sensitivity of the instrument in milli Ohms/ oC. Resistance (mῼ) Tempreture (oC) 210k 110 217k 89 221k 68 228k 47arrow_forwardA temperature measurement system is composed of a sensor and a readout device. The readout device has a claimed accuracy of 0.6 °C with a resolution of 0.1 °C. The sensor has an off-the-shelf limit of error of 0.5 °C. Estimate a design-stage uncertainty in the temperature indicated by this combination. A displacement transducer has the following specifications: Linearity error:±0.25% reading Drift:±0.05%/°C reading Sensitivity error: ±0.25% reading Excitation: 10-25 V Output: dc 0-5Vdc Range: 0-5 cm The transducer output is to be indicated on a voltme- ter having a stated accuracy of 0:1% reading with a resolution of 10 μV. The system is to be used at room temperature, which can vary by 10 °C. Estimate an uncertainty in a nominal displacement of 2 cm at the design stage. Assume 95% confidence.arrow_forwardIn an experiment designed to determine the value of a parameter n, the actual value that is determined during the experiment is ln n which is determined to be ln n = –1.17828±0.07023. What is the value of n? Make sure that you quote the answer to the correct number of significant figures.arrow_forward
- A pressure measurement system is stated as having the followingcharacteristics. Explain the significance of the terms:Range: 0 to 125 kPa and 0 to 2500 kPaAccuracy: ±1% of the displayed readingTemperature sensitivity: ±1% of the reading per o Carrow_forwardExplain the significance of the following information given in the specification of the followingtransducer,Thermocouple Sensitivity: nickel chromium/nickel aluminum thermocouple: 0.039 mV/ºC when thecold junction is at 0 ºC.arrow_forwardDiscuss the importance of measuring devices for Electric Motor and Internal Combustion Engine in terms of accuracy.arrow_forward
- An engineer sets the pressure in a supply tank usinga very accurate manometer as a guide and then reads theoutput of a 20-psig pressure gauge attached to the tank as10.2 psig. Sometime later she repeats the procedure andobtains values of 10.4 and 10.3 psig. Discuss the following gauge properties:(a)Precision(b)Accuracy(c)Resolution(d)RepeatabilityExpress these answers on a percentage of full-scale basis.arrow_forwardA thermometer having a time constant of 0.2 min is placed in a temperature bath, and after the thermometer comes to equilibrium with the bath, the temperature of the bath is increased linearly with time at a rate of l0 C/min. What is the difference between the indicated temperature and the bath temperature (a) 0.1 min, (b) 1.0 min after the change in temperature begins?arrow_forwardA thermometer is calibrated from 200degree centigrade to 300degree. the accuracy is specified as +-0.25%. Non Linearity is +-0.5% Hystersis is +-0.01% what is the maximum errorarrow_forward
arrow_back_ios
SEE MORE QUESTIONS
arrow_forward_ios
Recommended textbooks for you
- Principles of Heat Transfer (Activate Learning wi...Mechanical EngineeringISBN:9781305387102Author:Kreith, Frank; Manglik, Raj M.Publisher:Cengage Learning
Principles of Heat Transfer (Activate Learning wi...
Mechanical Engineering
ISBN:9781305387102
Author:Kreith, Frank; Manglik, Raj M.
Publisher:Cengage Learning
How to balance a see saw using moments example problem; Author: Engineer4Free;https://www.youtube.com/watch?v=d7tX37j-iHU;License: Standard Youtube License