The pH probe/meter uses following equations: Ecell = L + 0.0592 log a1 = L - 0.0592 pH Where L = L1 + EAg/AgCI + Easy= constants  L1 = - 0.0592 log a2  a1  = activity of analyte  solution            a2  = activity of internal solution   How will measured pH value  be affected vs “real” pH  if HCl in pH electrode,  became 0.15M instead of 0.1M ? a.impact can not be determined b.measured pH is higher than "real" pH. c.measured pH is lower than "real" pH. d.measured pH is same as "real" pH.

Biomedical Instrumentation Systems
1st Edition
ISBN:9781133478294
Author:Chatterjee
Publisher:Chatterjee
Chapter19: Biomedical Laboratory Instrumentation
Section: Chapter Questions
Problem 7P
icon
Related questions
Question

The pH probe/meter uses following equations:

Ecell = L + 0.0592 log a1 = L - 0.0592 pH

Where L = L1 + EAg/AgCI Easy= constants 

L1 = - 0.0592 log a2

 a1  = activity of analyte  solution            a2  = activity of internal solution 

 How will measured pH value  be affected vs “real” pH  if HCl in pH electrode,  became 0.15M instead of 0.1M ?

a.impact can not be determined

b.measured pH is higher than "real" pH.
c.measured pH is lower than "real" pH.
d.measured pH is same as "real" pH.
Expert Solution
steps

Step by step

Solved in 4 steps

Blurred answer
Similar questions
  • SEE MORE QUESTIONS
Recommended textbooks for you
Biomedical Instrumentation Systems
Biomedical Instrumentation Systems
Chemistry
ISBN:
9781133478294
Author:
Chatterjee
Publisher:
Cengage
Basic Clinical Laboratory Techniques 6E
Basic Clinical Laboratory Techniques 6E
Biology
ISBN:
9781133893943
Author:
ESTRIDGE
Publisher:
Cengage
Curren'S Math For Meds: Dosages & Sol
Curren'S Math For Meds: Dosages & Sol
Nursing
ISBN:
9781305143531
Author:
CURREN
Publisher:
Cengage