The pH probe/meter uses following equations: Ecell = L + 0.0592 log a1 = L - 0.0592 pH Where L = L1 + EAg/AgCI + Easy= constants L1 = - 0.0592 log a2 a1 = activity of analyte solution a2 = activity of internal solution How will measured pH value be affected vs “real” pH if HCl in pH electrode, became 0.15M instead of 0.1M ? a.impact can not be determined b.measured pH is higher than "real" pH. c.measured pH is lower than "real" pH. d.measured pH is same as "real" pH.
The pH probe/meter uses following equations: Ecell = L + 0.0592 log a1 = L - 0.0592 pH Where L = L1 + EAg/AgCI + Easy= constants L1 = - 0.0592 log a2 a1 = activity of analyte solution a2 = activity of internal solution How will measured pH value be affected vs “real” pH if HCl in pH electrode, became 0.15M instead of 0.1M ? a.impact can not be determined b.measured pH is higher than "real" pH. c.measured pH is lower than "real" pH. d.measured pH is same as "real" pH.
Biomedical Instrumentation Systems
1st Edition
ISBN:9781133478294
Author:Chatterjee
Publisher:Chatterjee
Chapter19: Biomedical Laboratory Instrumentation
Section: Chapter Questions
Problem 7P
Related questions
Question
The pH probe/meter uses following equations:
Ecell = L + 0.0592 log a1 = L - 0.0592 pH
Where L = L1 + EAg/AgCI + Easy= constants
L1 = - 0.0592 log a2
a1 = activity of analyte solution a2 = activity of internal solution
How will measured pH value be affected vs “real” pH if HCl in pH electrode, became 0.15M instead of 0.1M ?
a.impact can not be determined
b.measured pH is higher than "real" pH.
c.measured pH is lower than "real" pH.
d.measured pH is same as "real" pH.