Interpretation:
The type of mass spectrum used for molecular mass of compound and reason for using this mass spectrum is to be stated.
Concept introduction:
The mass spectrum of a compound is obtained by the ionization of the compound. The ionization can be done by using electron ionization or chemical ionization. The type in which the spectrum of compound obtained is simple and the process of the method is easy.
Answer to Problem 20.19QAP
The chemical ionization of the mass spectrum is used for molecular mass of a compound.
Explanation of Solution
The mass spectrum obtained by chemical ionization is used for molecular mass of compound because the molecular structure of the compound is disturbed in electron ionization. In electron ionization bombardment of electrons and photons is used to produce secondary ions. The energy consumed in electron ionization is high and the exact molecular structure is not obtained. If the molecular structure of the compound is different, then the mass spectrum obtained is also different and the molecular mass of the compound cannot be determined.
The chemical ionization process to obtain a spectrum of a compound is simple and consumes low energy. Thus, the chemical ionization of the mass spectrum is used for molecular mass of a compound.
(b)
Interpretation:
The type of mass spectrum used for the chemical structure of the compound and reason for using this is to be stated.
Concept introduction:
The chemical structure of the compound consists of the different orientation of atoms in a compound. The analysis of every atom is done to obtain a chemical structure of compound because a particular atom behaves differently with a different atom in the compound. The method used for the mass spectrum for the chemical structure of the compound should be such that the orientation of every atom can be analyzed separately.
Answer to Problem 20.19QAP
The electron ionization of the mass spectrum is used for the chemical structure of the compound.
Explanation of Solution
The electron ionization is used in mass spectrum for chemical structure of compound because every atom is analysed by ionizing differently and spectrum of chemical structure is very simple.
(c)
Interpretation:
The flight time for ion in TOF mass analyzer is to be stated.
Concept introduction:
In TOF mass analyzer positive ions are produced by bombarding the electrons or photons on molecules of a compound. These produced ions then accelerated in the drift tube due to the electric field applied. The ions in the tube are separated on the basis of their masses. The kinetic energies of all the ions in the tube are the same because all ions are under the effect of the same electric field. The velocities of ions are in an inverse of their masses.
Answer to Problem 20.19QAP
The flight time for ion in TOF mass analyzer is
Explanation of Solution
Write the expression of flight time for ion.
Here, the flight time is
Substitute
(d)
Interpretation:
The derivation of an equation for the differences in flight times as a function of two masses, charges and accelerating voltage is to be stated.
Concept introduction:
The ion in the tube of mass analyzer consists of kinetic energy and potential energy. The potential energy is due to charged electrons and kinetic energy due to the mass of ions. The difference in flight times is obtained by finding the flight time of each ion separately.
Answer to Problem 20.19QAP
The equation for the differences in flight times as a function of two masses, charges and accelerating voltage is
Explanation of Solution
Write the expression for potential energy of charged electrons
Here, charge of electron is
Write the expression of kinetic energy of ions
Here, mass of ions is
The potential energy of electrons and kinetic energy of ions is same in TOF mass analyzer. From Equation (II) and Equation (III).
Write the expression for flight time
Here, length of tube is
Substitute
Write the expression of the flight time for ion of mass to charge ratio
Write the expression of the flight time for ion of mass to charge ratio
Write the expression for difference in flight time of ions
Substitute
(e)
Interpretation:
The difference between flight time for ions of mass to charge ratios
Concept introduction:
The difference in flight time for ions depends on mass of ions, charge of electrons and accelerating voltage in the tube. It is given by the formula
Answer to Problem 20.19QAP
The difference between flight time for ions of mass to charge ratios
Explanation of Solution
Write the expression of flight time of ions
Here, the mass of ions are
Substitute
Write the expression of flight time of ions
Substitute
Write the expression for differences in flight time
Substitute
(f)
Interpretation:
The type of ionization source in tendom mass spectrometer and reason for using this is to be stated.
Concept introduction:
The mass spectrum is the plot between the relative abundance and mass to charge ratio. The mass spectrum is different for the same substance. The relative abundance represents the y-axis and the mass to charge ratio represents the x-axis. The heights of the peaks represent the relative intensity. The peak with the highest value in the plot is called the base peak. All the calculations are made with respect to this base peak. The structural information can be obtained by fragmentation of molecules in the sample.
Answer to Problem 20.19QAP
The electronic ionization is used to get the structural information of the compound because the single unpaired electron is obtained by electronic ionization.
Explanation of Solution
The fragmentation pattern is obtained by passing the vaporized organic sample from the ionization chamber. This vaporized organic sample is then bombarded by the stream of electrons and positive ions are obtained. This positive ion is called molecular ion and consists of a single unpaired electron. This single unpaired electron is obtained by the electronic ionization process. The peak in the mass spectra with value
(g)
Interpretation:
The types of mass spectra obtained from the MS/MS experiment is to be stated.
Concept introduction:
The mass spectrometry is a technique used to find the structure and chemical properties of different molecules. In the process of mass spectrometry, different molecules are converted into ions. The ions are separated by the mass to charge ratio and detected by relative abundance in the sample. The mass spectrum of ion is obtained by plotting the mass to charge ratio and relative abundance of the ion in a sample.
Answer to Problem 20.19QAP
The type of mass spectra obtained from MS/MS experiment by:
(1) Holding the first analyzer constant and scanning the second analyzer is product ion spectra.
(2) Scanning both analyzers with a small
(3) Scanning the first analyzer and holding the second analyzer constant is precursor ion spectra.
(4) Scanning the second mass analyzer for every mass-selected by the first mass analyzer is three-dimensional spectra.
Explanation of Solution
(1) The ion spectra obtained in the MS/MS experiment is very different from each other for any compound. The product ion spectra of the compound are obtained in the MS/MS experiment by holding the first analyzer constant and scanning the second analyzer.
(2).When the difference between mass to charge ratio to different ions is very small, by scanning both analyzers, then ions are detected by analyzing the neutral loss spectrum.
(3) The compounds with the same products in ions are identified by determining the concentration of ions in each compound. If ions are detected by scanning the first mass analyzer and keeping the second mass analyzer constant, then precursor ion spectra are obtained in the MS/MS experiment.
(4) If ions are detected by selecting ions in the first mass analyzer and scanned by the second mass analyzer, then the three-dimensional spectrum is obtained in the MS/MS experiment.
Want to see more full solutions like this?
Chapter 20 Solutions
Principles of Instrumental Analysis, 6th Edition
- A 10.00-g sample containing an analyte was transferred to a 250-mL volumetric flask and diluted to volume. When a 10.00-mL aliquot of the resulting solution was diluted to 25.00 mL it was found to give a signal of o.235 (arbitrary units). A second 10.00-mL aliquot was spiked with 10.00 mL of a 1.00- ppm standard solution of the analyte and diluted to 25.0o mL. The signal for the spiked sample was found to be o.502. Calculate the weight percent of analyte in the original sample.arrow_forwardWhich of the following compounds is most likely to give the below mass spectrum? Relative Intensity 100 80 8 90 20- 0 MS-NU-2643 25 Br O 50 75 100 125 150arrow_forwardChemistry Common Isotopes and Mass spectrum of element Mo are given below Most common isotopes: Mo-98; Mo-96 Mo-95; Mo-92; Mo-97 ; Mo-100; Mo-94 Calculate the average atomic mass of Mo How the MS determined the relative intensity (Scale of 100) and Isotope abundance (%)? Why absolute intensity was not used? Can detector sensitivity, ICPMS temperature influence the relative intensity? Explainarrow_forward
- An acceleration voltage of 15.0 kV is used to accelerate ions into the 1.48 m flight tube of a time-of-flight (TOF) mass spectrometer. The frequency at which spectra can be recorded is dependent on how long it takes the slowest ion to travel from the source to the detector. If you wish to scan up to m/z 13039, at what frequency (spectra/s) could spectra be recorded if a new spectrum begins each time the slowest ion reaches the detector? frequency: spectra/sarrow_forwardIn the interference experiment preformed, sodium ascorbate powder was added to water to prepare the stock concentration of the interfering substance. On the package label, the manufacturer states that every 1 gram (1,000mg) of sodium ascorbate powder contains 120mg of sodium and 880mg of ascorbate (vitamin C). How much sodium ascorbate must be weighed out (in milligrams) and added to 100 mL of water in order to have a final stock concentration of pure ascorbate equal to 250 mg/dL?arrow_forwardQuinine in a 1.939-g tablet was to be analyzed by dissolving it with 0.10 M HCl to a final volume of exactly 500 mL. This solution was diluted by adding 50.00 mL to a 100.0 mL flask and diluting to the mark. Afterwards the fluorescence intensity was found to be 394 in arbitrary units. Under identical experimental conditions, a standard 150.0-ppm had an intensity of 184.Quinine in a 1.939-g tablet was to be analyzed by dissolving it with 0.10 M HCl to a final volume of exactly 500 mL. This solution was diluted by adding 50.00 mL to a 100.0 mL flask and diluting to the mark. Afterwards the fluorescence intensity was found to be 394 in arbitrary units. Under identical experimental conditions, a standard 150.0-ppm had an intensity of 184. Calculate the number of milligrams in the original sample.arrow_forward
- You have extracted a new compound with a molar mass of 270 g/mol. A 10 mg/L of this substance in a 1-cm cuvette can transmit 60% of the original incident radiation intensity. Calculate (a) the absorbance and (b) the molar absorptivity of the new compound.arrow_forwardA 10 cm 3 air saturated Fricke dosimetry solution in a 1 cm diameter tube is irradiated for 10 min in a 6~ source of gamma rays. The optical density measured at 304 txm in a 1 cm light path at 30~ was 0.260 after the completion of irradiation and 0.003 before irradiation. (a) What is the total dose in gray absorbed by the solution?(b) If 10 cm 3 of methanol is irradiated in the same tube for the same 10min, what is the absorbed dose in gray. The Z/A for methanol is 0.562 and that for the Fricke dosimeter is 0.553. Explain any assumption you make to solve the problem.arrow_forwardstion Completion Status: Relative Abundance 100 90 80 70 60 50 40 30 20 10 0 20 21 22 Mass (amu) Consider the mass spectrum of neon above. It consists of an intense signal at mass number 20, an itty-bitty signal at 21, and a moderate signal at mass number 22. The heights of these signals are proportional to the number of counts at each mass number and, in turn, a reflection of the fractional abundance of the isotopes. The fractional abundance of each isotope is determined as follows: 1. The heights of all signals are measured with a metric ruler (in cm). 2. These heights are added together to give a total value. 3. The fractional abundance of an isotope is equal to the height of its signal divided by the sum of the heights of all the signals Which of the following best represents the height measurements for each isqope? (scaling so the mass spectrum box is about 10 cm tall before measuring with a metric ruler) 20Ne 9.25 cm, 21 Ne 0.02 cm, 22Ne 1.00 cm O20Ne 9 cm, 21 Ne 0.02 cm, 22Ne 1…arrow_forward
- As a forensic toxicologist, you have been given a task on determining amount of cadmium, mercury, chromium, copper and lead metal contents in medicinal herbs using flame atomic absorption spectroscopy (AAS). Hollow cathode lamp (HCL) was selected as radiation source for this analysis. The highest concentration came from lead element followed by chromium and copper, however cadmium and mercury were not detected in the sample. a) Based on your knowledge, conclude whether the above result is valid or not. b) If you are given an option of electrothermal or flame atomizer for above case study, determine which one is suitable and justify your answer.arrow_forwardHow does a CEM detector work?arrow_forwardA 10 μL aqueous sample containing 1 attogram of Rb+ was injected into a graphite furnace atomic absorption instrument. Calculate the concentration of Rb+ atoms in ppb. The molecular weight of Rb is 85.47 g/mol. density of water = 1.0 g/mL atto = 1 x 10-18 Amazingly, the low concentration of this sample can be detectedarrow_forward
- Principles of Instrumental AnalysisChemistryISBN:9781305577213Author:Douglas A. Skoog, F. James Holler, Stanley R. CrouchPublisher:Cengage LearningMacroscale and Microscale Organic ExperimentsChemistryISBN:9781305577190Author:Kenneth L. Williamson, Katherine M. MastersPublisher:Brooks Cole