Chapter 22, Problem 22.4P

(a)

Interpretation Introduction

Interpretation:

Reason behind the formation of more theoretical plates per meter than thick film has to be explained.

Concept Introduction:

Plate height can be calculated by van Deemter equation as follows:

    HA+Bu+(C)(u)

Here H is the plate height.

A term is eddy diffusion constant.

B term is longitudinal diffusion constant.

C term is mass transfer constant.

u is flow rate.

The term A is eddy diffusion constant and it is related to particle size of packing material and uniformity of the flow around the particles. It is independent of the flow rate.

The term B is longitudinal diffusion and it is related to diffusion of the analyte in mobile phase and on stationary phase.

The term C is mass transfer constant and it is related to linear velocity and square of the particle size.

Number of theoretical plates is calculated as follows:

    N=LH

Here,

N is number of theoretical plates.

L islength of the column.

H isplate height.

(a)

Expert Solution
Check Mark

Explanation of Solution

Particle size has a significant impact on the analyte band. For smaller sized particles at very low flow rate the second term in the van Deemter equation is high but other two terms are small so plate height is not high enough. At high flow rate, the third term cannot play significant role as it is proportional to the square of the particle size thus, the value of third term is not high enough. So the summation of all three terms is small. Hence plate height is low.

For lager sized particles at very low flow rate, second term in the van Deemter equation will be predominant, hence at low flow rate plate height is high. And at high flow rate, third term in the equation is predominant as it is proportional to the square of the particle size. So plate height is high.

Hence it can be concluded that in general pate height of thin films is greater than that of thick films because particle size in thin film is smaller than that of thick film.

As number of theoretical plates is inversely proportional to plate height hence number of theoretical plates is higher in case of this film.

(b)

Interpretation Introduction

Interpretation:

Mass of stationary phase in each column has to be calculated.

Concept Introduction:

Formula to calculate volume of stationary phase is calculated as follows:

    Volumeofstationaryphase=[(π)(Plateheight)(Columndiameter)(Thicknessofstationaryphase)]

(b)

Expert Solution
Check Mark

Explanation of Solution

For narrow bore column mass of stationary phase is 15.7ng and for wide bore column mass of stationary phase is 5.55ng.

For narrow bore open tubular gas chromatography plate height is calculated as follows:

    Plateheight=LengthofcolumnNumberofplates=1m5000=2.0×104m

The volume of stationary phase is calculated as follows:

    Volumeofstationaryphase=[(π)(Plateheight)(Columndiameter)(Thicknessofstationaryphase)]=(π)(2×104m)(0.250×103m)(0.1×106m)=15.7×1015m3(106mL1m3)=15.7×109mL

Mass of stationary phase is calculated as follows:

    Massofstationaryphase=(Density)(Volume)=(1.00gmL1)(15.7×109mL)=15.7×109g=15.7ng

For wide bore open tubular gas chromatography plate height is calculated as follows:

    Plateheight=LengthofcolumnNumberofplates=1m1500=6.6667×104m

The volume of stationary phase is calculated as follows:

    Volumeofstationaryphase=((π)(Plateheight)(Columndiameter)(Thicknessofstationaryphase))=(π)(6.667×104m)(0.53×106m)(5×106m)=5.55×1015m3(106mL1m3)=5.55×109mL

Mass of stationary phase is calculated as follows:

    Massofstationaryphase=(Density)(Volume)=(1.00g/mL)(5.55×109mL)=5.55×109g=5.55ng

For narrow bore column mass of stationary phase is 15.7ng and for wide bore column mass of stationary phase is 5.55ng.

(c)

Interpretation Introduction

Interpretation:

Mass of analyte can be injected has to be calculated.

Concept Introduction:

Refer to part (b).

(c)

Expert Solution
Check Mark

Answer to Problem 22.4P

For narrow bore column 0.157ng analyte can be injected and for wide bore column 0.0555ng analyte can be injected.

Explanation of Solution

For narrow bore open tubular gas chromatography plate height is calculated as follows:

    Plateheight=LengthofcolumnNumberofplates=1m5000=2.0×104m

The volume of stationary phase is calculated as follows:

    Volumeofstationaryphase=(π)(Plateheight)(Columndiameter)(Thicknessofstationaryphase)=(π)(2×104m)(0.250×103m)(0.1×106m)=15.7×1015m3=15.7×109mL

Mass of stationary phase is calculated as follows:

    Massofstationaryphase=(Density)(Volume)=(1.00gmL1)(15.7×109mL)=15.7×109g=15.7ng

Mass of analyte can be injected is calculated as follows:

    Massofanalyteinjected=(Massofstationaryphase)(1100)=(15.7ng)(1100)=0.157ng

For wide bore open tubular gas chromatography plate height is calculated as follows:

    Plateheight=LengthofcolumnNumberofplates=1m1500=6.6667×104m

The volume of stationary phase is calculated as follows:

    Volumeofstationaryphase=(π)(Plateheight)(Columndiameter)(Thicknessofstationaryphase)=(π)(6.667×104m)(0.53×106m)(5×106m)=5.55×1015m3=5.55×109mL

Mass of stationary phase is calculated as follows:

    Massofstationaryphase=(Density)(Volume)=(1.00gmL1)(5.55×109mL)=5.55×109g=5.55ng

Mass of analyte can be injected is calculated as follows:

    Massofanalyteinjected=(Massofstationaryphase)(1100)=(5.55ng)(1100)=0.0555ng

For narrow bore column 0.157ng analyte can be injected and for wide bore column 0.0555ng analyte can be injected.

Want to see more full solutions like this?

Subscribe now to access step-by-step solutions to millions of textbook problems written by subject matter experts!
Knowledge Booster
Background pattern image
Recommended textbooks for you
Text book image
Chemistry
Chemistry
ISBN:9781305957404
Author:Steven S. Zumdahl, Susan A. Zumdahl, Donald J. DeCoste
Publisher:Cengage Learning
Text book image
Chemistry
Chemistry
ISBN:9781259911156
Author:Raymond Chang Dr., Jason Overby Professor
Publisher:McGraw-Hill Education
Text book image
Principles of Instrumental Analysis
Chemistry
ISBN:9781305577213
Author:Douglas A. Skoog, F. James Holler, Stanley R. Crouch
Publisher:Cengage Learning
Text book image
Organic Chemistry
Chemistry
ISBN:9780078021558
Author:Janice Gorzynski Smith Dr.
Publisher:McGraw-Hill Education
Text book image
Chemistry: Principles and Reactions
Chemistry
ISBN:9781305079373
Author:William L. Masterton, Cecile N. Hurley
Publisher:Cengage Learning
Text book image
Elementary Principles of Chemical Processes, Bind...
Chemistry
ISBN:9781118431221
Author:Richard M. Felder, Ronald W. Rousseau, Lisa G. Bullard
Publisher:WILEY