Food Dyes Analysis of Commercial Products
.pdf
keyboard_arrow_up
School
University of South Florida *
*We aren’t endorsed by this school
Course
2045L
Subject
Chemistry
Date
Apr 3, 2024
Type
Pages
3
Uploaded by ChefRainBear7
Food Dyes Analysis of Commercial Products
CHM2045L.017S24
Mar 21, 2024
Taylor Tran
Annabelle Martin
Sophia Bates
Maadh Alshaaili
Objective:
The purpose of this experiment is to understand and apply the principles of spectrophotometry
for chemical analysis. We will also need to be able to explain the interaction of UV-visible
radiation and molecules in commercial products. Creating and understanding a calibration curve
is also required. Alongside the calibration curve, understanding the electromagnetic spectrum,
light wavelength and color, UV-Visible absorption spectroscopy, spectrophotometer,
concentration of solutions, and dilution will also be needed.
Materials:
●
25mL volumetric flask
●
100mL beaker
●
7mL Graduated Pipettes
●
Red Dye 40
●
Blue Dye 1
●
Pipette pumps
●
Cuvettes
●
Pasco Scientific Vis-Spectrophotometer
Chemicals/Hazards:
Chemical Name
Chemical
Formula
Molecular
Formula Weight
Physical
Appearance
Specific Hazards
Dihydrogen
monoxide
H
2
O
18.01528 g/mol
Clear liquid
N/A
Red #40
C
18
H
14
N
2
Na
2
O
8
S
2
498.44 g/mol
Dark Red
Powder
May cause eye
irritation.
May cause
gastrointestinal
discomfort.
May cause
irritation to the
respiratory tract.
May cause
irritation to skin.
Blue #1
C
37
H
34
N
2
O
9
S
3
Na
2
792.86 g/mol
Dark violet
Powder
May cause eye
irritation.
May cause
gastrointestinal
discomfort.
May cause
irritation to the
respiratory tract.
May cause
irritation to skin.
Methods and Procedures:
Creating a Calibration Curve Using a Spectrophotometer:
1- We will begin by calibrating the spectrophotometer using the black (dark) cuvette and a
regular cuvette filled with DI water to calibrate the dark and light values respectively.
2- Now we start by finding the theoretical and experimental values of lambda max (max
wavelength) by observing the color of the sample and then placing it in the
spectrophotometer to obtain a reading.
3- Using a parallel dilution method we will dilute the original sample into 4 more samples going
down linearly in concentration.
We will start with the stock solution and dilute it into ratios of (4/10), (3/10),(2/10), and
(1/10) by simply making a ratio of stock solution and DI water.
4- Then we will measure the absorbance value of each sample (including the original) making
sure to note the calculated concentration associated with the absorbance levels.
5- Using this data we can plot the obtained absorption values against the calculated concentration
values to obtain a calibration curve and the equation represented by the curve.
6- Repeat this procedure for each dye used.
Your preview ends here
Eager to read complete document? Join bartleby learn and gain access to the full version
- Access to all documents
- Unlimited textbook solutions
- 24/7 expert homework help
Related Questions
Carbon monoxide is another common poison found in the forensic laboratory. Explain how UV-visible spectrophotometry is applied to quantify carbon monoxide in blood samples.
arrow_forward
Calculate the molar extinction coefficient (e) of a molecule at a particular wavelength given that a 0.000764 M solution of the
molecule produced a fluorescence intensity (I) of 1177 when the irradiance of the beam incident to the sample (Po) was 821 and
the length of the medium (b) was 1.12 cm. The proportionality constant k' for this molecule is 0.319.
€ =
×10
TOOLS
M-1.cm-1
arrow_forward
Please explain the answer.
Plotting the ratio of atomic emission of Pb to
the atomic emission of yttrium (internal
standard) vs the concentration of Pb in
contaminated water, the linear regression
equation from 10-point calibration was found
to be y=0.114x-1.8872. What would be the
average concentration of Pb in contaminated
water if three aliquot of a sample of 10 folds'
diluted solution of contaminated water
provided the ratio of emission of Pb to that of
yttrium as: 3.92, 3.86, and 3.87?
а. 3.89 ppb
b. 38.9 ppb
с. 357 ppb
d. 35.7 ppb
arrow_forward
The luminescent first-order decay of a certain organic molecule yields the following data
t/s
1
2
4
10
I
100
60.65
36.79
22.31
13.53
8.21
0.67
where I is the relative intensity. Calculate the mean lifetime t for the process.
1.59
X S
Is it fluorescence or phosphorescence?
O fluorescence
arrow_forward
5
arrow_forward
Given C6H12O2 chemical formula. chemical shift , intergration, multiplicity and interpretation needed.
arrow_forward
Flame Atomic Absorption Spectroscopy (FAAS) method is selected due to its wide applicability for metal ion analysis in drinking water samples.
Provide appropriate discussion to determine the concentration of selected heavy metals (Cu and Cr) in the unknown tap water sample.
arrow_forward
Suppose the difference in rest energy of the neutron and proton were Qn = (mn − mp)c2 = x MeV instead of Qn = 1.29MeV, with all other physical parameters unchanged. Estimate Ymax, the maximum possible mass fraction in 4He, assuming that all available neutrons are incorporated into 4He nuclei. Hint: Compare the new rest energy of the neutron to the electron and consider how that would impact the decay of neutrons. Will free neutrons decay at all in this scenario? If the neutron decay is negligible, then the ratio of protons to neutrons to protons will remain the same as it was at the time of freeze-out (for a given Qn).
Value: x=0.158 MeV
arrow_forward
using the following information to construct a valid hypothesis addressing the effect of pH and temperature on the hydrolysis of starch.
Per bench, prepare the stock amylase enzyme solution by adding distilled water to the powdered enzyme in the flask , each bench will share one flaks of enzyme solution.
per groups, turn on the spectrometer and set the wavelength to 560nm.allow the instruments to warm up for at least 10 minutes.
per group, set up 17 cuvettes in a Styrofoam rack. Add 3 drops of iodine (IKI) to each coveter.
per group, prepare the “black” cuvette by rapidly adding 3 mL of maltose standard solution to one of The prepared cuvettes containing IKI. Rapid addition of the maltose will mix the maltose and IKI
arrow_forward
What is the purpose of a calibration curve in analytical chemistry?
arrow_forward
A company is producing a powder which can be reconstituted as a
healthy drink for the elderly. Analysis of the powder for calcium involved
using the following procedure.
4 g of powder was reconstituted to 150 ml with water. Thirty ml of 24 %
TCA was added to 30 ml of this solution. The mixture was allowed to
stand for 15 mins and then filtered. The filtrate was analysed for calcium
using atomic absorption spectroscopy. From the standard curve it was
found that the reading for the analysed sample was 24mg/litre.
Calculate the concentration of calcium in the powder (mg/kg).
arrow_forward
Instrumentation of IR spectroscopy? Please answer at your own words.
arrow_forward
The space immediately surrounding stars, the circumstellar space, is significantly warmer because stars are very intense black-body emitters with temperatures of several thousand kelvin. Discuss how such factors as cloud temperature, particle density, and particle velocity may affect the rotational spectrum of CO in an interstellar cloud. What new features in the spectrum of CO can be observed in gas ejected from and still near a star with temperatures of about 1000 K, relative to gas in a cloud with temperature of about 10 K? Explain how these features may be used to distinguish between circumstellar and interstellar material on the basis of the rotational spectrum of CO.
arrow_forward
For a molecule, the phosphorescence lifetime without quenching is 1.0 s while with a quencher population of 10-2 M the phosphorescence lifetime is 0.10 s. Calculate the rate constant for phosphorescence assuming that kisc >> (kF + kec + kic). Calculate the rate constant for external conversion of the triplet state (kec’), the quenching rate constant for the triplet state (kq’), and the Stern-Volmer quenching constant. Assume that intersystem crossing from T1 to So is negligible.
arrow_forward
Chemistry
0.1237 (±0.0001) g of sodium carbonate (Na2CO3) required 22.46 (±0.02) cm3 of aqueous HCl for complete neutralisation. Determine the concentration of the HCl in mol dm–3 together with its uncertainty in mol dm–3 to an appropriate number of digits. You may assume the error in the molar mass of Na2CO3 is negligible.
arrow_forward
The IR spectrum of 2-methyl-1-butanol (H;CCH,CH(CH;)
CH,OH) and tert-butyl methyl ether [(CH;);COCH3] are
shown below. Assign cach spectrum to the correct com-
pound and identify the frequencies and the functional
groups used to support your assignment.
λ (μm)
6 7 8 9 10 11 12 1314 15
2.5
100
3
5
90
80
70
60
50
40
30
20
10
4000 3600 3200 2800 2400 2000 1800 1600 1400 1200 1000
800
600
400
Frequency (cm-1)
A (µm)
7 8 9 10 11 12 13 14 15
2.5
100
3
5
20
4
6
90
80
70
60
50
40
30
20
10
4000 3600 3200 2800 2400 2000 1800 1600 1400 1200 1000
800
600
400
Frequency (cm¬!)
(%) Transmission
(%) Transmission
20
arrow_forward
An atomic absorption method for the determination of the amount of iron present in used jet engine oil was found from pooling 30 triplicate analyses to have a standard deviation s = 3.6 mg Fe/mL. If s is a good estimate of σ, calculate the 95 and 99% confidence intervals for the result 18.5 mg Fe/mL if it was based on (a) a single analysis, (b) the mean of two analyses, and (c) the mean of four analyses.
arrow_forward
(c) Comment on the truth, or otherwise, of the following statement. "Fast time-scale
flash photolysis can be used to determine rate constants for collisional relaxation
(quenching) processes of singlet and triplet states. Radiative decay rate constants
(Einstein A coefficients) can be determined from such measurements, and can also be
calculated from absorption coefficients."
arrow_forward
A kinetic experiment is carried out in a flow tube where the flow rate is 25 m s¯¹. The detector is a
spectrophotometer, and absorbances are measured at various distances along the tube. The
reactant has a molar absorption coefficient of 1180 mol-¹dm³ cm¹ at 420 nm. The absorbances are
measured using a 1 cm optical cell. Use this information, and that in the following table, to calculate
[reactant] as a function of time.
Then plot a graph of [reactant] versus time and comment on its shape.
May plot using excel.
distance
cm
A
2.0
0.575
4.0
0.457
6.0
0.346
8.0
0.232
10.0
0.114
arrow_forward
CPU/RAM
ProctorU e Proctor
e Proctoru
o.com/collab/ui/session/join/43918c0bf9cb421ca7b98b0075f5f239
Firefox Ex
SPECTROSCOPY
1. If the concentration of solute in a solution is high, the Absorbance will
be high or low and why?
2. Was the spectroscopy experiment qualitative or quantitative analysis?
3. If we have 12M HCI and we need 3 L of 1M HCI, how much 12 M
should we use?
4. What does Molarity mean?
5. When must safety goggles be worn?
6. If you make a mistake recording data, what must you do?
spec hw.pptx (2/3)
..%
An
arrow_forward
The Na* concentration in a blood serum sample was determined using the method of
standard addition and atomic spectroscopy. 25.00 mL aliquots of serum were pipetted into
five 50.00 mL volumetric flasks into which 2.640 M NaCl standard was added as shown in
the table below. After dilution to volume, the emission of each solution was measured and
used to plot a calibration graph.
Flask
1
23
4
5
Vol. of
standard (mL)
0
1.00
2.00
3.00
4.00
Emission
signal
3.13
5.40
7.89
10.30
12.48
Analytical signal (emission)
14
42 10
12
4
2
0
0
0.05
y = 44.697x + 3.12
R² = 0.9995
0.1
0.15
Concentration of added analyte (M)
0.2
a) Using the plotted calibration graph, calculate the Na* concentration in the blood serum
sample.
b) Provide a possible reason why the serum sample might have been analysed using this
analytical procedure.
0.25
arrow_forward
17. Advection is represented in the advection-dispersion equation
by the term
+v Əc/ax
+KC
-v ac/ax
-μС
arrow_forward
Infrared Spectroscopy (IR) is very useful for determining the functional groups present
in an organic molecule. For each of the spectra below, complete the following:
• Build the molecule in Spartan and describe the motion of the molecule at each key
absorption frequency
• Identify key absorption frequencies in the spectrum and identify the type of
stretch/bend/wag/etc. (i.e. O-H)
•
Match the spectra from the provided structures
•
Identify the main functional groups in the molecule
OH
A
OH
H
B
E
F
H
N
NH2
arrow_forward
SEE MORE QUESTIONS
Recommended textbooks for you
Principles of Modern Chemistry
Chemistry
ISBN:9781305079113
Author:David W. Oxtoby, H. Pat Gillis, Laurie J. Butler
Publisher:Cengage Learning
Chemistry & Chemical Reactivity
Chemistry
ISBN:9781337399074
Author:John C. Kotz, Paul M. Treichel, John Townsend, David Treichel
Publisher:Cengage Learning
Chemistry & Chemical Reactivity
Chemistry
ISBN:9781133949640
Author:John C. Kotz, Paul M. Treichel, John Townsend, David Treichel
Publisher:Cengage Learning
Principles of Instrumental Analysis
Chemistry
ISBN:9781305577213
Author:Douglas A. Skoog, F. James Holler, Stanley R. Crouch
Publisher:Cengage Learning
Chemistry for Engineering Students
Chemistry
ISBN:9781285199023
Author:Lawrence S. Brown, Tom Holme
Publisher:Cengage Learning
Related Questions
- Carbon monoxide is another common poison found in the forensic laboratory. Explain how UV-visible spectrophotometry is applied to quantify carbon monoxide in blood samples.arrow_forwardCalculate the molar extinction coefficient (e) of a molecule at a particular wavelength given that a 0.000764 M solution of the molecule produced a fluorescence intensity (I) of 1177 when the irradiance of the beam incident to the sample (Po) was 821 and the length of the medium (b) was 1.12 cm. The proportionality constant k' for this molecule is 0.319. € = ×10 TOOLS M-1.cm-1arrow_forwardPlease explain the answer. Plotting the ratio of atomic emission of Pb to the atomic emission of yttrium (internal standard) vs the concentration of Pb in contaminated water, the linear regression equation from 10-point calibration was found to be y=0.114x-1.8872. What would be the average concentration of Pb in contaminated water if three aliquot of a sample of 10 folds' diluted solution of contaminated water provided the ratio of emission of Pb to that of yttrium as: 3.92, 3.86, and 3.87? а. 3.89 ppb b. 38.9 ppb с. 357 ppb d. 35.7 ppbarrow_forward
- The luminescent first-order decay of a certain organic molecule yields the following data t/s 1 2 4 10 I 100 60.65 36.79 22.31 13.53 8.21 0.67 where I is the relative intensity. Calculate the mean lifetime t for the process. 1.59 X S Is it fluorescence or phosphorescence? O fluorescencearrow_forward5arrow_forwardGiven C6H12O2 chemical formula. chemical shift , intergration, multiplicity and interpretation needed.arrow_forward
- Flame Atomic Absorption Spectroscopy (FAAS) method is selected due to its wide applicability for metal ion analysis in drinking water samples. Provide appropriate discussion to determine the concentration of selected heavy metals (Cu and Cr) in the unknown tap water sample.arrow_forwardSuppose the difference in rest energy of the neutron and proton were Qn = (mn − mp)c2 = x MeV instead of Qn = 1.29MeV, with all other physical parameters unchanged. Estimate Ymax, the maximum possible mass fraction in 4He, assuming that all available neutrons are incorporated into 4He nuclei. Hint: Compare the new rest energy of the neutron to the electron and consider how that would impact the decay of neutrons. Will free neutrons decay at all in this scenario? If the neutron decay is negligible, then the ratio of protons to neutrons to protons will remain the same as it was at the time of freeze-out (for a given Qn). Value: x=0.158 MeVarrow_forwardusing the following information to construct a valid hypothesis addressing the effect of pH and temperature on the hydrolysis of starch. Per bench, prepare the stock amylase enzyme solution by adding distilled water to the powdered enzyme in the flask , each bench will share one flaks of enzyme solution. per groups, turn on the spectrometer and set the wavelength to 560nm.allow the instruments to warm up for at least 10 minutes. per group, set up 17 cuvettes in a Styrofoam rack. Add 3 drops of iodine (IKI) to each coveter. per group, prepare the “black” cuvette by rapidly adding 3 mL of maltose standard solution to one of The prepared cuvettes containing IKI. Rapid addition of the maltose will mix the maltose and IKIarrow_forward
- What is the purpose of a calibration curve in analytical chemistry?arrow_forwardA company is producing a powder which can be reconstituted as a healthy drink for the elderly. Analysis of the powder for calcium involved using the following procedure. 4 g of powder was reconstituted to 150 ml with water. Thirty ml of 24 % TCA was added to 30 ml of this solution. The mixture was allowed to stand for 15 mins and then filtered. The filtrate was analysed for calcium using atomic absorption spectroscopy. From the standard curve it was found that the reading for the analysed sample was 24mg/litre. Calculate the concentration of calcium in the powder (mg/kg).arrow_forwardInstrumentation of IR spectroscopy? Please answer at your own words.arrow_forward
arrow_back_ios
SEE MORE QUESTIONS
arrow_forward_ios
Recommended textbooks for you
- Principles of Modern ChemistryChemistryISBN:9781305079113Author:David W. Oxtoby, H. Pat Gillis, Laurie J. ButlerPublisher:Cengage LearningChemistry & Chemical ReactivityChemistryISBN:9781337399074Author:John C. Kotz, Paul M. Treichel, John Townsend, David TreichelPublisher:Cengage LearningChemistry & Chemical ReactivityChemistryISBN:9781133949640Author:John C. Kotz, Paul M. Treichel, John Townsend, David TreichelPublisher:Cengage Learning
- Principles of Instrumental AnalysisChemistryISBN:9781305577213Author:Douglas A. Skoog, F. James Holler, Stanley R. CrouchPublisher:Cengage LearningChemistry for Engineering StudentsChemistryISBN:9781285199023Author:Lawrence S. Brown, Tom HolmePublisher:Cengage Learning
Principles of Modern Chemistry
Chemistry
ISBN:9781305079113
Author:David W. Oxtoby, H. Pat Gillis, Laurie J. Butler
Publisher:Cengage Learning
Chemistry & Chemical Reactivity
Chemistry
ISBN:9781337399074
Author:John C. Kotz, Paul M. Treichel, John Townsend, David Treichel
Publisher:Cengage Learning
Chemistry & Chemical Reactivity
Chemistry
ISBN:9781133949640
Author:John C. Kotz, Paul M. Treichel, John Townsend, David Treichel
Publisher:Cengage Learning
Principles of Instrumental Analysis
Chemistry
ISBN:9781305577213
Author:Douglas A. Skoog, F. James Holler, Stanley R. Crouch
Publisher:Cengage Learning
Chemistry for Engineering Students
Chemistry
ISBN:9781285199023
Author:Lawrence S. Brown, Tom Holme
Publisher:Cengage Learning