Chapter 16, Problem 16.10QAP
Interpretation Introduction

(a)

Interpretation:

The length of the mirror derives in a spectrometer at resolution of 0.010cm1 is to be calculated.

Concept introduction:

The FTIR-stands for the Fourier Transform Infrared Spectroscopy. Generally, this technique is used to determine the organic, inorganic and polymeric materials.

The length of the mirror derives in a spectrometer is the half of the distance between the mirror derives. The distance between the mirror derives is reciprocal of the resolution of the FTIR spectrometer.

Interpretation Introduction

(b)

Interpretation:

The length of the mirror derives in a spectrometer at resolution of 0.50cm1 is to be calculated.

Concept introduction:

The FTIR-stands for the Fourier Transform Infrared Spectroscopy. Generally, this technique is used to determine the organic, inorganic and polymeric materials.

The length of the mirror derives in a spectrometer is the half of the distance between the mirror derives. The distance between the mirror derives is reciprocal of the resolution of the FTIR spectrometer.

Interpretation Introduction

(c)

Interpretation:

The length of the mirror derives in a spectrometer at resolution of 2.0cm1 is to be calculated.

Concept introduction:

The FTIR-stands for the Fourier Transform Infrared Spectroscopy. Generally, this technique is used to determine the organic, inorganic and polymeric materials.

The length of the mirror derives in a spectrometer is the half of the distance between the mirror derives. The distance between the mirror derives is reciprocal of the resolution of the FTIR spectrometer.

Blurred answer
Students have asked these similar questions
What length of mirror drive in a Michelson interferometer is required to produce a resolution sufficient to separate (a) IR peaks at 600.7 and 600.4 cm-1?
Q6. (a) Describe the differences between a linear TOF mass spectrometer and an orthogonal TOF instrument? What are the advantages and limitations of these two geometries? * (b) What function does the ICP torch have in ICPMS? (c) Why has ICPMS become an important and widely used analytical method?
Answer the questions regarding a single Fourier transform infrared spectroscopy (FTIR) experiment. The interferogram was sampled at a retardation interval (8) of 0.0001297 cm. What wavenumber range is possible with a sample retardation interval (8) of 0.0001297 cm? Aỹ = cm- For a spectrum acquired in 8-bit space (256 data points sampled) for ô = -A to d = +A , determine the value of maximum retardation (A) and the resolution (R). A = cm R = cm-1 What is the interferometer mirror velocity if each data point required 267 us (tintt) to obtain? Vint = cm/s
Knowledge Booster
Background pattern image
Chemistry
Learn more about
Need a deep-dive on the concept behind this application? Look no further. Learn more about this topic, chemistry and related others by exploring similar questions and additional content below.
Similar questions
Recommended textbooks for you
Text book image
Principles of Instrumental Analysis
Chemistry
ISBN:9781305577213
Author:Douglas A. Skoog, F. James Holler, Stanley R. Crouch
Publisher:Cengage Learning
Text book image
Fundamentals Of Analytical Chemistry
Chemistry
ISBN:9781285640686
Author:Skoog
Publisher:Cengage