INSTRUMENTAL ANALYSIS-ACCESS >CUSTOM<
INSTRUMENTAL ANALYSIS-ACCESS >CUSTOM<
7th Edition
ISBN: 9781337783439
Author: Skoog
Publisher: CENGAGE C
Question
100%
Book Icon
Chapter 26, Problem 26.16QAP
Interpretation Introduction

(a)

Interpretation:

The resolution for species B and C from the given data should be determined.

Concept introduction:

The resolution of the column is defined as the separation of two species of the column.

Expert Solution
Check Mark

Answer to Problem 26.16QAP

The resolution is 2.01.

Explanation of Solution

Given:

INSTRUMENTAL ANALYSIS-ACCESS >CUSTOM<, Chapter 26, Problem 26.16QAP

The expression of resolution of the column is:

RS=2[(tR)C(tR)B]WC+WB ....... (I)

Here, the retention time of species C is (tR)C, the retention time of species B is (tR)B, the width time of the peak base of the species C is WC and the width time of the peak base of the species B is WB.

Substitute 15.7min for (tR)B, 13.3min for (tR)C, 1.07min for WC and 1.32min for WD in Equation (I).

RS=2[(15.7min)(13.3min)](1.07min)+(1.32min)=2(2.4min)(2.39min)=2.01

Thus, the resolution is 2.01.

Interpretation Introduction

(b)

Interpretation:

The selectivity factor for species B and C from the given data should be determined.

Concept introduction:

The resolution of the column is defined as the separation of two species of the column.

Expert Solution
Check Mark

Answer to Problem 26.16QAP

The selectivity factor is 1.24.

Explanation of Solution

The expression of selectivity factor is:

aC,B=(tR)CtM(tR)BtM ........ (II)

Here, the non-retained retention time is tM.

Substitute 15.7min for (tR)C, 13.3min for (tR)B and 3.1min for tM in Equation (II).

aC,B=(15.7min)(3.1min)(13.3min)(3.1min)=12.6min10.2min=1.24

Thus, the selectivity factor is 1.24.

Interpretation Introduction

(c)

Interpretation:

The length of column necessary to separate B and C species with a resolution of 2.5 should be determined.

Concept introduction:

The resolution of the column is defined as the separation of two species of the column.

Expert Solution
Check Mark

Answer to Problem 26.16QAP

The length of column necessary to separate the two species with a resolution of 2.5 is 37.98cm.

Explanation of Solution

The expression of length of the column is:

L=N2H ....... (III)

Here, the number of plates needed to separate the two species is N and the height of the plate is H.

The expression of relation of the resolution and number of plates is:

(RS)(RS)2=N1N2N2=(RS)2N1(RS)N2={(RS)2N1(RS)}2N2=N1(RS)22(RS)2

Here, the number of plates needed is N1 and the given resolution is (RS)2.

Substitute N1(RS)22(RS)2 for N2 in Equation (III).

L={N1(RS)22(RS)2}H ......... (IV)

Substitute 2.01 for (RS), 2.5 for (RS)2, 2455 for N1 and 0.01cm for H in Equation (IV).

L={(2455)(2.5)2(2.01)2}(0.01cm)=(3797.86)(0.01cm)=37.98cm

Thus, the length of column necessary to separate the two species with a resolution of 2.5 is 37.98cm.

Interpretation Introduction

(d)

Interpretation:

The time required to separate B and C species on the column of part c should be determined.

Concept introduction:

The resolution of the column is defined as the separation of two species of the column.

Expert Solution
Check Mark

Answer to Problem 26.16QAP

The time required to separate the two species on the column of part c is 24.29min.

Explanation of Solution

The expression of the relation of time required to separate the two species on the column is:

(tR)C(tR)=(Rs)2(RS)12 ........ (V)

Here, the given resolution is (RS)1 and time required to separate the two species on the column is tR.

Substitute 15.7min for (tR)C, 2.01 for RS and 2.5 for (RS)1 in Equation (V).

(15.7min)(tR)=(2.01)2(2.5)2tR=(15.7min)(2.5)2(2.01)2tR=24.29min

Thus, the time required to separate the two species on the column of part c is 24.29min.

Want to see more full solutions like this?

Subscribe now to access step-by-step solutions to millions of textbook problems written by subject matter experts!
Students have asked these similar questions
What is the role of specific visualizing agents in a Thin layer Chromatography? Please shortly write at your own words. Answer should be to the point (5-6 lines maximum ).
In your own words answer the following With reference to the Van Deemter equation, explain thoroughly whether separation may be expected to improve in an Liquid solid chromatography  LSC column with a 5 μm particle size or the one with a 20 μm particle size
(b) For column 1(i) What are the unadjusted retention times for components A and B?
Knowledge Booster
Background pattern image
Similar questions
SEE MORE QUESTIONS
Recommended textbooks for you
Text book image
Fundamentals Of Analytical Chemistry
Chemistry
ISBN:9781285640686
Author:Skoog
Publisher:Cengage
Text book image
Principles of Instrumental Analysis
Chemistry
ISBN:9781305577213
Author:Douglas A. Skoog, F. James Holler, Stanley R. Crouch
Publisher:Cengage Learning
Text book image
Appl Of Ms Excel In Analytical Chemistry
Chemistry
ISBN:9781285686691
Author:Crouch
Publisher:Cengage