Question
Book Icon
Chapter 15, Problem 15.8QAP
Interpretation Introduction

(a)

Interpretation:

A working curve from the data should be constructed.

Concept introduction:

Standard addition method is used to analyze complex samples which matrix effect is significant. One common approach is standard addition technique is spiking of sample. In this method, one or more increments of a standard solution is added to the several aliquots of sample containing same volume. Each solution is then diluted to a fixed volume and experiment is proceeded.

Interpretation Introduction

(b)

Interpretation:

A linear least-squares equation for the data should be determined.

Concept introduction:

Syy=(yiy¯)2Sxx=(xix¯)2Sxy=(xix¯)(yiy¯)

The slope of the line, m=SxySxx

The intercept, b=y¯mx¯

Interpretation Introduction

(c)

Interpretation:

The standard deviation of the slope and intercept and the standard deviation about regression for the curve should be determined.

Concept introduction:

Standard deviation about regression, sr=Syym2SxxN2

N − number of points used.

The standard deviation of the slope, sm=sr2Sxx

The standard deviation of the intercept, sb=srxi2Nxi2(xi)2

Interpretation Introduction

(d)

Interpretation:

The concentration of Zn2+ in the sample should be calculated.

Concept introduction:

Standard addition method is used to analyze complex samples which matrix effect is significant. One common approach is standard addition technique is spiking of sample. In this method, one or more increments of a standard solution is added to the several aliquots of sample containing same volume. Each solution is then diluted to a fixed volume and experiment is proceeded.

Interpretation Introduction

(e)

Interpretation:

The standard deviation for the result in part (d) should be calculated.

Concept introduction:

Standard deviation of the results obtained from standard addition method = sc=Cx(smm)2+(sbb)2

M − number of replicates

N- number of calibration points.

Blurred answer
Students have asked these similar questions
The following volumes of a solution containing 1.10 ppm Zn2+ were pipetted into separatory funnels each containing 5.00 mL of an unknown zinc solution: 0.00, 1.00, 4.00, 8.00 and 12.00. Each was extracted with three 5 mL aliquots of CCl4 containing an excess of 8-hydroxyquinoline. The extracts were then diluted to 25.0 mL and their fluorescence measured with a fluorometer. The results for the standard addition calibration were as follows: Volume Std. Zn2+ / mL [Zn2+] / ppm Fluorometer Reading 0.00   6.12 1.00   7.41 4.00   11.16 8.00   15.68 12   20.64 a) Determine the concentration on Zn2+ ions in each of the standard solutions coming from the coming from the 1.10 ppm Zn2+ solutions, Please show your working
Exactly 10.00-mL aliquots of a solution containing phenobarbital were measured into 50.00-mL volumetric flasks and made basic with KOH. The following volumes of a standard solution of phenobarbital containing 2.000μg/mL of phenobarbital were then introduced into each flask and the mixture was diluted to volume: 0.000, 0.500, 1.00, 1.50, 2.00 mL. The fluorescence of each of these solutions was measured with a fluorimeter, which gave values of 3.26, 4.80, 6.41, 8.02, 9.56, respectively. a. plot the data.  b. derive a least squares equation for the data plotted in (a). c. find the concentration of phenobarbital from the equation in (b). d. calculate a standard deviation for the concentration obtained in (c).
One dietary supplement capsule weighing 4.90 was ground into a fine powder. Two portions of the solid, both weighing 0.200 g, were dissolved in dilute acid, and transferred to 50-mL volumetric flask. To one of these, 5.00 mL of 40.0 ppm manganese ion was added, then both flasks were diluted to the mark with distilled water. When aspirated into the flame of an atomic absorption spectrometer set at a manganese absorption wavelength, the absorbances of the unknown and unknown plus standard were 0.300 and 0.650, respectively. What is the type of calibration method used? a) Standard Addition b) Internal Standard c) External Standard
Knowledge Booster
Background pattern image
Similar questions
SEE MORE QUESTIONS
Recommended textbooks for you
Text book image
Fundamentals Of Analytical Chemistry
Chemistry
ISBN:9781285640686
Author:Skoog
Publisher:Cengage
Text book image
Principles of Instrumental Analysis
Chemistry
ISBN:9781305577213
Author:Douglas A. Skoog, F. James Holler, Stanley R. Crouch
Publisher:Cengage Learning