Chapter 8 Challange Questions
.docx
keyboard_arrow_up
School
Cypress College *
*We aren’t endorsed by this school
Course
RADT-150
Subject
Chemistry
Date
Apr 3, 2024
Type
docx
Pages
2
Uploaded by ccstudent
Challenge Questions
Yolanda Cuevas
Chapter 8: X-ray Emission
1)
Define or otherwise identify the following:
a)
Inherent filtration - The filtration provided by the window and oil bath of the x-ray tube.
b)
The unit of x-ray quantity
- Milliroentgen (mR) or milligray (mGy
a
).
f)
Half-value layer - The thickness of the filter or other absorbing material that will reduce intensity to half its original value.
g)
Wedge filter - A compensating filter designed to shape the useful x-ray beam to the anatomic structure under examination so that the exposure of the image receptor will be more uniform.
h)
The unit of x-ray quality - Identified numerically by half-value layer; also influenced by kVp and the added filtration in the useful beam.
j) X-ray intensity - X-ray intensity is another term for x-ray quantity, measured in mR or mGy.
3) An abdominal radiograph taken at 84 kVp, 150 mAs result in patient radiation exposure of 6.5 mGy a
. The image
is too light and is repeated at 84 kVp, 250 mAs. What is the new radiation exposure? 10.83 mGy
a
(1083) mR
4) An image of the lateral skull taken at 68 kVp, 20 mAs has sufficient optical density but too much contrast. If the
kVp is increased to 78 kVp, what should be the new mAs? 10 mAs
5)
The chest radiograph taken at 180 cm SID results in an exposure of 120 uGy
a
. What would the exposure be if the same radiographic factors were used at 100 cm SID? 389 uGy
a
(38.9 mR)
6)
The following data were obtained with a fluoroscopic x-ray tube operated at 80 kVp: The exposure levels were measured 50cm above the patient couch with aluminum absorbers positioned on the surface of the couch. Estimate the HVL through visual inspection of the data; then plot the data & determine the precise value of the HVL.
Added mm Al uGy
a
None
650
1
480
3
300
5
210
7
160
9
130
One half of 650 uGy
a
(65 mR) is 325 uGy
a
(32.5 mR), which falls between 1 and 3 mm Al. A
plot shows the correct answer to be 2.8 mm Al.
10)
An extremity is radiographed at 60 kVp, 10 mAs, resulting in an x-ray intensity of 280 uGy
a
. If the technique is changed to 55 kVp, 10 mAs, what is the resultant x-ray intensity? 235 uGy
a
(23.5 mR).
11) What is the square law, and how is it used? The square law involves compensating for a change in SID by changing the mAs by the factor SID
2
. When SID is increased, the mAs must be increased by SID
2
to maintain constant exposure of the image receptor.
12)
What is the primary purpose of x-ray beam filtration? To remove low-energy x-rays, which only increase patient
dose without contributing to the image.
13)
The kVp is reduced from 78 to 68 kVp. What, if anything, should be done with mAs to maintain exposure of the image receptor constant? Double the mAs.
14) What is the relationship between x-ray quantity and mAs? Directly proportional. Double mAs; double x-ray quantity.
15)
Define half-value layer. The thickness of absorber necessary to reduce the x-ray beam intensity to half.
16)
List the two ways an x-ray beam can be shifted to a higher average energy. Increase the kVp or add filtration.
17)
Why is aluminum used for x-ray beam filtration? Because it is efficient at removing low-energy x-rays, and because it is readily available, inexpensive, and easily shaped into filters.
18)
Describe the use of a wedge filter during radiography of a foot. The thicker part of the wedge should be positioned toward the toes in order to equalize optical density.
19)
Does adding filtration to the x-ray beam affect the quantity of x-rays reaching the image receptor? Yes, the quantity is reduced with an increase in filtration.
20)
Increasing Effect on X-ray Quality
Effect of X-ray Quantity
mAs
None
Increase
kVp
Increase
Increase
Distance
None
Reduce
Filtration
Increase
Reduce
Your preview ends here
Eager to read complete document? Join bartleby learn and gain access to the full version
- Access to all documents
- Unlimited textbook solutions
- 24/7 expert homework help
Related Questions
Chemistry
tabulate between HLC and plasma radiation sources.
arrow_forward
Wavelength (nm)
a. Which part of the visible light spectrum is most useful (most absorbed) by
chlorophyll b?
b. Which pigment do you think was best at absorbing the red light that was
shown on the leaves when using a red-light filter in the experiment?
Explain.
c. Do you think a greenhouse with green and yellow lights would be very
successful in terms of plant growth? Explain.
100%
n(US)
home
end
insert
prt sc
F10
F12
DII
F6
FB
F9
F11
F7
F3
F4
F5
&
%23
2$
4.
5
7
8.
Y
U
F
G
H
arrow_forward
A bcps.schoology.com/common-assessment-delivery/start/4865831278?action=Donresume&submissionld%3514311155
aSchoology
S Schoology
If the student increased the rate at which drops were added to the pan, this would increase the
v of the waves that were generated.
Figure 1
1
2
4.
6.
7
9.
Co
arrow_forward
Please complete Table 2 and show the calculations for protein concentration and the concentration of the undiluted sample.
a). Using Excel software, enter data from Table 1 to plot a standard curve (corrected average absorbance on Y-axis and protein concentrations on X-axis). The standard curve must contain correct labels on both axes.
b). From the standard curve, obtain a best fit line and equation.
c). Use the best fit equation and data from Table 2 to find the concentrations of proteins in sample A2-D2 and E1. [onto your plotted standard curve, indicate where A595 of A2-D2 and E1 are located]
d). Calculate the concentrations of proteins in undiluted samples A to E.
Thank you very much.
arrow_forward
True or False.
1. The absorbance of a sample is equal to the negative logarithm of the transmittance.
2. Winkler method of dissolve oxygen determination is an example of iodometry.
3. A diazo compound is produced when nitrite is reacted with sulfanilamide in acidic solution.
4. Atomic Absorption Spectroscopy can be used to analyze the functional groups of organic substances.
5. A diode array spectrometer use an absorption or interference filters for wavelength selection and a photoelectric device for measuring radiant power.
6. The oxidation state for the manganese in the brown solid which precipitate in Winkler method is +3
7. In Winkler method, the iodine produced is oxidized during sodium thiosulfate titration to iodide ion.
8. When the pH of the solution is adjusted to pH greater than 12, the EDTA reacts with magnesium ions only, because calcium precipitates as calcium hydroxide.
arrow_forward
10. For the flame shown in Figure 9-3, calculate the relative intensity of the 766.5-nm emission
line for potassium in the flame center and at the following heights above the orifice (assume
no ionization and use the 2.0 cm for comparison). (a) 3.0 cm (b) 5.0 cm
1700
1750
5.0
1600
1800
1400
1830
4.0
1858
1863
3.0
2.0
1830
1800
1.0
1700
1.5
1.0 0.5
0.5 1.0
1.5
cm
cm
Burner tip
Distance above orifice, cm
arrow_forward
Here is the protocol for a UV-Vis spectrophotometer to detect water and chlorine-carbon.
1.Dissolve the water and chlorine-carbon compounds in a solvent, such as water.
2.Prepare a standard solution of known concentration that is similar to the sample being measured.
3.Calibrate the spectrophotometer using the standard solution.
4.Measure the absorbance of the sample using the spectrophotometer.
5.Calculate the concentration of the compounds in the sample using the calibration curve obtained from the standard solution.
How is the spectrophotometer calibrated with standard solutions? When is the blank solution placed in the spectrophotmeter?
arrow_forward
1. Give the general steps in the use of a spectrophotometer.
2. Define the following:
a. ) Optical Density
b.) Transmittance
c.) Beer's Law
arrow_forward
Timer
Course Contents » ... » Data, Statistics, and Graphing » Graphing Absorbance (Optical Density) versus Molar Concentration
The following table contains some data showing the effect of molar concentration of a solution on the absorbance (optical density) reading from a spectrophotometer.
No
↑ Does the solution in your graph follow Beer's Law?
Concentration (M)
Submit Answer Incorrect. Tries 37/99 Previous Tries
0.13
0.28
0.43
0.58
0.73
0.88
1.03
1.18
1.33
1.48
What is the concentration of a solution that has an Optical Density of 0.21? (Give units in your answer).
0.21M
You are correct.
Your receipt no. is 155-7520 ?
What is the concentration of a solution that has an Optical Density of 0.91? (Give units in your answer).
0.81 M
Previous Tries
Optical Density
To assist in graphing, right click on this excel spreadsheet file, save it to your computer, and open it in Excel. Enter the above data in the appropriate spaces in the spreadsheet. Select the graph that is created, and
print…
arrow_forward
Pre-Lab Questions- To be completed BEFORE lab as an entry slip. These questions are based on
content presented in the lab procedure and in lecture.
1. Consider the electromagnetic spectrum. (Reference Site)
-- Increasing Frequency (v)
10² 10%
10
10² 10%
Trays
1044
10
10-
10
UV
30*
10
IR
Visible spectrum
10²
10*
30
Microwave FM AM
Radio waves
10° 10²
10 H 10²
Increasing Wavelength (2)→
10
Increasing Wavelength in m
a. What type of radiation has the longest wavelength?
10 10² 10 voto
b. What type of radiation has the highest energy?
Long radio waves
c. Looking at the emission line spectrum for hydrogen in Figure 1. Which colored line
corresponds to the emission of the highest-energy photon?
arrow_forward
This question uses the same data as the previous question. The calibration curve is a graph of absorbance versus concentration. Absorbance is a dimensionless quantity. The slope of the calibration curve is m = 2.24 x 10^4 M^-1. Find the concentration detection limit.
arrow_forward
5
What would the effect of increasing sample pressure (while maintaining a fixed cell length) have on an FTIR spectrum. By ”cell” I mean the chamber holding the gas.
arrow_forward
the spectra is of a drug either cannabinoid, opiate, MDMA etc, please explain the spectra and what drug could it be
arrow_forward
This expresses the direct proportionality between absorbance and concentration.
(A) Beer-Lambert's Law
B) Dalton's Law
Charles' Law
Gay-Lussac's Law
arrow_forward
As a forensic toxicologist, you have been given a task on determining amount of cadmium, mercury, chromium, copper and lead metal contents in medicinal herbs using flame atomic absorption spectroscopy (AAS). Hollow cathode lamp (HCL) was selected as radiation source for this analysis. The highest concentration came from lead element followed by chromium and copper, however cadmium and mercury were not detected in the sample.
a) Based on your knowledge, conclude whether the above result is valid or not.
b) If you are given an option of electrothermal or flame atomizer for above case study, determine which one is suitable and justify your answer.
arrow_forward
What is the purpose of reference beam in the double-beam spectrophotometer shown in Figure 20-1?
Select all the correct answers.
a)To correct for changes in source light intensity as a function of time
b)to correct for changes in intensity of the light source as a function of wavelength
c)to correct of scattering of light off the surface of the sample cuvette
d)to correct for absorbance of the solvent
e)to correct for changes in detector signal as a function of wavelength
arrow_forward
What causes exposure or closeness in humans: [Explain in detail and thenjustify why]a) to an IR lamp?
arrow_forward
The given question was incorrectly answered, please identify what the errors made were. Then correctly answer and explain the given the question.
arrow_forward
Hi please answer all and below the answe place the URL from where you got the answers
thank you
arrow_forward
1. State the difference between transmittance, absorbance, molar absorptivity, and
fluorescence. Which one(s) is proportional to analyte concentration?
arrow_forward
what do you think the UV-vis spectrum of a sample that was not clear (there was particulare in the sample) would look like? Would the particulates in solution affect the quality of the spectrum? What if the cuvette has a fingerprint on it? Explain your answer.
arrow_forward
Name of Experiment:
Fluorescence Spectroscopy
Q1: Purpose of
experiment
Q2: Introduction
arrow_forward
PLEASE ANSWER ASAP
arrow_forward
Macmi
After you have made use of your blank and set the wavelength, transfer 3 mL of your reaction flask solution into one of the
8 remaining cuvettes. This will be measured in the spectrophotometer and serve as your time zero reading.
What should you notice about the solution in this cuvette?
arrow_forward
Determine the standard concentration of A through E and then the absorbance. I just need clarification on if I add together the two ml values given.
arrow_forward
Help interpreting IR.
arrow_forward
SEE MORE QUESTIONS
Recommended textbooks for you
Principles of Instrumental Analysis
Chemistry
ISBN:9781305577213
Author:Douglas A. Skoog, F. James Holler, Stanley R. Crouch
Publisher:Cengage Learning
Related Questions
- Chemistry tabulate between HLC and plasma radiation sources.arrow_forwardWavelength (nm) a. Which part of the visible light spectrum is most useful (most absorbed) by chlorophyll b? b. Which pigment do you think was best at absorbing the red light that was shown on the leaves when using a red-light filter in the experiment? Explain. c. Do you think a greenhouse with green and yellow lights would be very successful in terms of plant growth? Explain. 100% n(US) home end insert prt sc F10 F12 DII F6 FB F9 F11 F7 F3 F4 F5 & %23 2$ 4. 5 7 8. Y U F G Harrow_forwardA bcps.schoology.com/common-assessment-delivery/start/4865831278?action=Donresume&submissionld%3514311155 aSchoology S Schoology If the student increased the rate at which drops were added to the pan, this would increase the v of the waves that were generated. Figure 1 1 2 4. 6. 7 9. Coarrow_forward
- Please complete Table 2 and show the calculations for protein concentration and the concentration of the undiluted sample. a). Using Excel software, enter data from Table 1 to plot a standard curve (corrected average absorbance on Y-axis and protein concentrations on X-axis). The standard curve must contain correct labels on both axes. b). From the standard curve, obtain a best fit line and equation. c). Use the best fit equation and data from Table 2 to find the concentrations of proteins in sample A2-D2 and E1. [onto your plotted standard curve, indicate where A595 of A2-D2 and E1 are located] d). Calculate the concentrations of proteins in undiluted samples A to E. Thank you very much.arrow_forwardTrue or False. 1. The absorbance of a sample is equal to the negative logarithm of the transmittance. 2. Winkler method of dissolve oxygen determination is an example of iodometry. 3. A diazo compound is produced when nitrite is reacted with sulfanilamide in acidic solution. 4. Atomic Absorption Spectroscopy can be used to analyze the functional groups of organic substances. 5. A diode array spectrometer use an absorption or interference filters for wavelength selection and a photoelectric device for measuring radiant power. 6. The oxidation state for the manganese in the brown solid which precipitate in Winkler method is +3 7. In Winkler method, the iodine produced is oxidized during sodium thiosulfate titration to iodide ion. 8. When the pH of the solution is adjusted to pH greater than 12, the EDTA reacts with magnesium ions only, because calcium precipitates as calcium hydroxide.arrow_forward10. For the flame shown in Figure 9-3, calculate the relative intensity of the 766.5-nm emission line for potassium in the flame center and at the following heights above the orifice (assume no ionization and use the 2.0 cm for comparison). (a) 3.0 cm (b) 5.0 cm 1700 1750 5.0 1600 1800 1400 1830 4.0 1858 1863 3.0 2.0 1830 1800 1.0 1700 1.5 1.0 0.5 0.5 1.0 1.5 cm cm Burner tip Distance above orifice, cmarrow_forward
- Here is the protocol for a UV-Vis spectrophotometer to detect water and chlorine-carbon. 1.Dissolve the water and chlorine-carbon compounds in a solvent, such as water. 2.Prepare a standard solution of known concentration that is similar to the sample being measured. 3.Calibrate the spectrophotometer using the standard solution. 4.Measure the absorbance of the sample using the spectrophotometer. 5.Calculate the concentration of the compounds in the sample using the calibration curve obtained from the standard solution. How is the spectrophotometer calibrated with standard solutions? When is the blank solution placed in the spectrophotmeter?arrow_forward1. Give the general steps in the use of a spectrophotometer. 2. Define the following: a. ) Optical Density b.) Transmittance c.) Beer's Lawarrow_forwardTimer Course Contents » ... » Data, Statistics, and Graphing » Graphing Absorbance (Optical Density) versus Molar Concentration The following table contains some data showing the effect of molar concentration of a solution on the absorbance (optical density) reading from a spectrophotometer. No ↑ Does the solution in your graph follow Beer's Law? Concentration (M) Submit Answer Incorrect. Tries 37/99 Previous Tries 0.13 0.28 0.43 0.58 0.73 0.88 1.03 1.18 1.33 1.48 What is the concentration of a solution that has an Optical Density of 0.21? (Give units in your answer). 0.21M You are correct. Your receipt no. is 155-7520 ? What is the concentration of a solution that has an Optical Density of 0.91? (Give units in your answer). 0.81 M Previous Tries Optical Density To assist in graphing, right click on this excel spreadsheet file, save it to your computer, and open it in Excel. Enter the above data in the appropriate spaces in the spreadsheet. Select the graph that is created, and print…arrow_forward
- Pre-Lab Questions- To be completed BEFORE lab as an entry slip. These questions are based on content presented in the lab procedure and in lecture. 1. Consider the electromagnetic spectrum. (Reference Site) -- Increasing Frequency (v) 10² 10% 10 10² 10% Trays 1044 10 10- 10 UV 30* 10 IR Visible spectrum 10² 10* 30 Microwave FM AM Radio waves 10° 10² 10 H 10² Increasing Wavelength (2)→ 10 Increasing Wavelength in m a. What type of radiation has the longest wavelength? 10 10² 10 voto b. What type of radiation has the highest energy? Long radio waves c. Looking at the emission line spectrum for hydrogen in Figure 1. Which colored line corresponds to the emission of the highest-energy photon?arrow_forwardThis question uses the same data as the previous question. The calibration curve is a graph of absorbance versus concentration. Absorbance is a dimensionless quantity. The slope of the calibration curve is m = 2.24 x 10^4 M^-1. Find the concentration detection limit.arrow_forward5 What would the effect of increasing sample pressure (while maintaining a fixed cell length) have on an FTIR spectrum. By ”cell” I mean the chamber holding the gas.arrow_forward
arrow_back_ios
SEE MORE QUESTIONS
arrow_forward_ios
Recommended textbooks for you
- Principles of Instrumental AnalysisChemistryISBN:9781305577213Author:Douglas A. Skoog, F. James Holler, Stanley R. CrouchPublisher:Cengage Learning
Principles of Instrumental Analysis
Chemistry
ISBN:9781305577213
Author:Douglas A. Skoog, F. James Holler, Stanley R. Crouch
Publisher:Cengage Learning