Chapter 8 Challange Questions
docx
School
Cypress College *
*We aren’t endorsed by this school
Course
RADT-150
Subject
Chemistry
Date
Apr 3, 2024
Type
docx
Pages
2
Uploaded by ccstudent
Challenge Questions
Yolanda Cuevas
Chapter 8: X-ray Emission
1)
Define or otherwise identify the following:
a)
Inherent filtration - The filtration provided by the window and oil bath of the x-ray tube.
b)
The unit of x-ray quantity
- Milliroentgen (mR) or milligray (mGy
a
).
f)
Half-value layer - The thickness of the filter or other absorbing material that will reduce intensity to half its original value.
g)
Wedge filter - A compensating filter designed to shape the useful x-ray beam to the anatomic structure under examination so that the exposure of the image receptor will be more uniform.
h)
The unit of x-ray quality - Identified numerically by half-value layer; also influenced by kVp and the added filtration in the useful beam.
j) X-ray intensity - X-ray intensity is another term for x-ray quantity, measured in mR or mGy.
3) An abdominal radiograph taken at 84 kVp, 150 mAs result in patient radiation exposure of 6.5 mGy a
. The image
is too light and is repeated at 84 kVp, 250 mAs. What is the new radiation exposure? 10.83 mGy
a
(1083) mR
4) An image of the lateral skull taken at 68 kVp, 20 mAs has sufficient optical density but too much contrast. If the
kVp is increased to 78 kVp, what should be the new mAs? 10 mAs
5)
The chest radiograph taken at 180 cm SID results in an exposure of 120 uGy
a
. What would the exposure be if the same radiographic factors were used at 100 cm SID? 389 uGy
a
(38.9 mR)
6)
The following data were obtained with a fluoroscopic x-ray tube operated at 80 kVp: The exposure levels were measured 50cm above the patient couch with aluminum absorbers positioned on the surface of the couch. Estimate the HVL through visual inspection of the data; then plot the data & determine the precise value of the HVL.
Added mm Al uGy
a
None
650
1
480
3
300
5
210
7
160
9
130
One half of 650 uGy
a
(65 mR) is 325 uGy
a
(32.5 mR), which falls between 1 and 3 mm Al. A
plot shows the correct answer to be 2.8 mm Al.
10)
An extremity is radiographed at 60 kVp, 10 mAs, resulting in an x-ray intensity of 280 uGy
a
. If the technique is changed to 55 kVp, 10 mAs, what is the resultant x-ray intensity? 235 uGy
a
(23.5 mR).
11) What is the square law, and how is it used? The square law involves compensating for a change in SID by changing the mAs by the factor SID
2
. When SID is increased, the mAs must be increased by SID
2
to maintain constant exposure of the image receptor.
12)
What is the primary purpose of x-ray beam filtration? To remove low-energy x-rays, which only increase patient
dose without contributing to the image.
13)
The kVp is reduced from 78 to 68 kVp. What, if anything, should be done with mAs to maintain exposure of the image receptor constant? Double the mAs.
14) What is the relationship between x-ray quantity and mAs? Directly proportional. Double mAs; double x-ray quantity.
15)
Define half-value layer. The thickness of absorber necessary to reduce the x-ray beam intensity to half.
16)
List the two ways an x-ray beam can be shifted to a higher average energy. Increase the kVp or add filtration.
17)
Why is aluminum used for x-ray beam filtration? Because it is efficient at removing low-energy x-rays, and because it is readily available, inexpensive, and easily shaped into filters.
18)
Describe the use of a wedge filter during radiography of a foot. The thicker part of the wedge should be positioned toward the toes in order to equalize optical density.
19)
Does adding filtration to the x-ray beam affect the quantity of x-rays reaching the image receptor? Yes, the quantity is reduced with an increase in filtration.
20)
Increasing Effect on X-ray Quality
Effect of X-ray Quantity
mAs
None
Increase
kVp
Increase
Increase
Distance
None
Reduce
Filtration
Increase
Reduce
Your preview ends here
Eager to read complete document? Join bartleby learn and gain access to the full version
- Access to all documents
- Unlimited textbook solutions
- 24/7 expert homework help
Related Documents
Related Questions
A bcps.schoology.com/common-assessment-delivery/start/4865831278?action=Donresume&submissionld%3514311155
aSchoology
S Schoology
If the student increased the rate at which drops were added to the pan, this would increase the
v of the waves that were generated.
Figure 1
1
2
4.
6.
7
9.
Co
arrow_forward
Please complete Table 2 and show the calculations for protein concentration and the concentration of the undiluted sample.
a). Using Excel software, enter data from Table 1 to plot a standard curve (corrected average absorbance on Y-axis and protein concentrations on X-axis). The standard curve must contain correct labels on both axes.
b). From the standard curve, obtain a best fit line and equation.
c). Use the best fit equation and data from Table 2 to find the concentrations of proteins in sample A2-D2 and E1. [onto your plotted standard curve, indicate where A595 of A2-D2 and E1 are located]
d). Calculate the concentrations of proteins in undiluted samples A to E.
Thank you very much.
arrow_forward
Here is the protocol for a UV-Vis spectrophotometer to detect water and chlorine-carbon.
1.Dissolve the water and chlorine-carbon compounds in a solvent, such as water.
2.Prepare a standard solution of known concentration that is similar to the sample being measured.
3.Calibrate the spectrophotometer using the standard solution.
4.Measure the absorbance of the sample using the spectrophotometer.
5.Calculate the concentration of the compounds in the sample using the calibration curve obtained from the standard solution.
How is the spectrophotometer calibrated with standard solutions? When is the blank solution placed in the spectrophotmeter?
arrow_forward
1. Give the general steps in the use of a spectrophotometer.
2. Define the following:
a. ) Optical Density
b.) Transmittance
c.) Beer's Law
arrow_forward
Timer
Course Contents » ... » Data, Statistics, and Graphing » Graphing Absorbance (Optical Density) versus Molar Concentration
The following table contains some data showing the effect of molar concentration of a solution on the absorbance (optical density) reading from a spectrophotometer.
No
↑ Does the solution in your graph follow Beer's Law?
Concentration (M)
Submit Answer Incorrect. Tries 37/99 Previous Tries
0.13
0.28
0.43
0.58
0.73
0.88
1.03
1.18
1.33
1.48
What is the concentration of a solution that has an Optical Density of 0.21? (Give units in your answer).
0.21M
You are correct.
Your receipt no. is 155-7520 ?
What is the concentration of a solution that has an Optical Density of 0.91? (Give units in your answer).
0.81 M
Previous Tries
Optical Density
To assist in graphing, right click on this excel spreadsheet file, save it to your computer, and open it in Excel. Enter the above data in the appropriate spaces in the spreadsheet. Select the graph that is created, and
print…
arrow_forward
detector
Answer Bank
emission monochromator
excitation monochromator
light source
sample cell
^ex
A
em
It is typically a grating, prism, or filter that selects a narrow wavelength band of radiation and passes it
to the sample.
It holds the sample and has a defined pathlength.
It is typically a photomultiplier tube that generates an electric signal when struck by photons.
It is positioned 90° to the incident light and selects a narrow wavelength band of radiation to pass to
arrow_forward
This question uses the same data as the previous question. The calibration curve is a graph of absorbance versus concentration. Absorbance is a dimensionless quantity. The slope of the calibration curve is m = 2.24 x 10^4 M^-1. Find the concentration detection limit.
arrow_forward
What causes exposure or closeness in humans: [Explain in detail and thenjustify why]a) to an IR lamp?
arrow_forward
The given question was incorrectly answered, please identify what the errors made were. Then correctly answer and explain the given the question.
arrow_forward
Hi please answer all and below the answe place the URL from where you got the answers
thank you
arrow_forward
NAME:
LAB SECTION:
EXPERIMENT 21: Spectroscopy of Dyes
POSTLAB EXERCISE
1. AUV/V is a spectrometer used to measure the absorbance of a solution of hemoglobin bound to oxy-
gen at 590 nm. The molar absorptivity was determined to be 8.57 × 104 M-'cm-. The cuvette used
had a 1.00-cm sample pathlength. The measured absorbance was ), 140. What was the concentra-
tion of hemoglobin in the solution?
2. What would be the absorbance of the solution in question 1 if the pathlength were 2.00 cm?
nohntooote
196
arrow_forward
what do you think the UV-vis spectrum of a sample that was not clear (there was particulare in the sample) would look like? Would the particulates in solution affect the quality of the spectrum? What if the cuvette has a fingerprint on it? Explain your answer.
arrow_forward
PLEASE ANSWER ASAP
arrow_forward
Macmi
After you have made use of your blank and set the wavelength, transfer 3 mL of your reaction flask solution into one of the
8 remaining cuvettes. This will be measured in the spectrophotometer and serve as your time zero reading.
What should you notice about the solution in this cuvette?
arrow_forward
Determine the standard concentration of A through E and then the absorbance. I just need clarification on if I add together the two ml values given.
arrow_forward
9) Sketch a diagram of an atomic absorption spectrometer. Label the nebulizer, light source
(be specific), monochromator, and detector. Describe the purpose of each portion of the
instrument.
arrow_forward
SEE MORE QUESTIONS
Recommended textbooks for you

Principles of Instrumental Analysis
Chemistry
ISBN:9781305577213
Author:Douglas A. Skoog, F. James Holler, Stanley R. Crouch
Publisher:Cengage Learning
Related Questions
- A bcps.schoology.com/common-assessment-delivery/start/4865831278?action=Donresume&submissionld%3514311155 aSchoology S Schoology If the student increased the rate at which drops were added to the pan, this would increase the v of the waves that were generated. Figure 1 1 2 4. 6. 7 9. Coarrow_forwardPlease complete Table 2 and show the calculations for protein concentration and the concentration of the undiluted sample. a). Using Excel software, enter data from Table 1 to plot a standard curve (corrected average absorbance on Y-axis and protein concentrations on X-axis). The standard curve must contain correct labels on both axes. b). From the standard curve, obtain a best fit line and equation. c). Use the best fit equation and data from Table 2 to find the concentrations of proteins in sample A2-D2 and E1. [onto your plotted standard curve, indicate where A595 of A2-D2 and E1 are located] d). Calculate the concentrations of proteins in undiluted samples A to E. Thank you very much.arrow_forwardHere is the protocol for a UV-Vis spectrophotometer to detect water and chlorine-carbon. 1.Dissolve the water and chlorine-carbon compounds in a solvent, such as water. 2.Prepare a standard solution of known concentration that is similar to the sample being measured. 3.Calibrate the spectrophotometer using the standard solution. 4.Measure the absorbance of the sample using the spectrophotometer. 5.Calculate the concentration of the compounds in the sample using the calibration curve obtained from the standard solution. How is the spectrophotometer calibrated with standard solutions? When is the blank solution placed in the spectrophotmeter?arrow_forward
- 1. Give the general steps in the use of a spectrophotometer. 2. Define the following: a. ) Optical Density b.) Transmittance c.) Beer's Lawarrow_forwardTimer Course Contents » ... » Data, Statistics, and Graphing » Graphing Absorbance (Optical Density) versus Molar Concentration The following table contains some data showing the effect of molar concentration of a solution on the absorbance (optical density) reading from a spectrophotometer. No ↑ Does the solution in your graph follow Beer's Law? Concentration (M) Submit Answer Incorrect. Tries 37/99 Previous Tries 0.13 0.28 0.43 0.58 0.73 0.88 1.03 1.18 1.33 1.48 What is the concentration of a solution that has an Optical Density of 0.21? (Give units in your answer). 0.21M You are correct. Your receipt no. is 155-7520 ? What is the concentration of a solution that has an Optical Density of 0.91? (Give units in your answer). 0.81 M Previous Tries Optical Density To assist in graphing, right click on this excel spreadsheet file, save it to your computer, and open it in Excel. Enter the above data in the appropriate spaces in the spreadsheet. Select the graph that is created, and print…arrow_forwarddetector Answer Bank emission monochromator excitation monochromator light source sample cell ^ex A em It is typically a grating, prism, or filter that selects a narrow wavelength band of radiation and passes it to the sample. It holds the sample and has a defined pathlength. It is typically a photomultiplier tube that generates an electric signal when struck by photons. It is positioned 90° to the incident light and selects a narrow wavelength band of radiation to pass toarrow_forward
- This question uses the same data as the previous question. The calibration curve is a graph of absorbance versus concentration. Absorbance is a dimensionless quantity. The slope of the calibration curve is m = 2.24 x 10^4 M^-1. Find the concentration detection limit.arrow_forwardWhat causes exposure or closeness in humans: [Explain in detail and thenjustify why]a) to an IR lamp?arrow_forwardThe given question was incorrectly answered, please identify what the errors made were. Then correctly answer and explain the given the question.arrow_forward
- Hi please answer all and below the answe place the URL from where you got the answers thank youarrow_forwardNAME: LAB SECTION: EXPERIMENT 21: Spectroscopy of Dyes POSTLAB EXERCISE 1. AUV/V is a spectrometer used to measure the absorbance of a solution of hemoglobin bound to oxy- gen at 590 nm. The molar absorptivity was determined to be 8.57 × 104 M-'cm-. The cuvette used had a 1.00-cm sample pathlength. The measured absorbance was ), 140. What was the concentra- tion of hemoglobin in the solution? 2. What would be the absorbance of the solution in question 1 if the pathlength were 2.00 cm? nohntooote 196arrow_forwardwhat do you think the UV-vis spectrum of a sample that was not clear (there was particulare in the sample) would look like? Would the particulates in solution affect the quality of the spectrum? What if the cuvette has a fingerprint on it? Explain your answer.arrow_forward
arrow_back_ios
SEE MORE QUESTIONS
arrow_forward_ios
Recommended textbooks for you
- Principles of Instrumental AnalysisChemistryISBN:9781305577213Author:Douglas A. Skoog, F. James Holler, Stanley R. CrouchPublisher:Cengage Learning

Principles of Instrumental Analysis
Chemistry
ISBN:9781305577213
Author:Douglas A. Skoog, F. James Holler, Stanley R. Crouch
Publisher:Cengage Learning